High voltage – Checkline LS-9-12000 User Manual
Page 2
1.0 Introduction
Congratulations on your purchase of a Checkline® LS-9-12000 Linear Stroboscope.
It uses the principle of stroboscopic lighting to create an illusion of stopped motion,
allowing detailed inspection of objects moving at high speeds.
NOTE: When using a stroboscope, it is important to remember that while objects may
appear to become motionless, they are, in fact, still moving. Exercise proper care to
prevent injury.
Please read this instruction manual thoroughly before operating your stroboscope. If you
have any questions not answered by this manual, call your local supplier, or visit our
web chat interface at www.checkline.com for full product support.
1.1 Package Contents
LS-9-12000 Stroboscope is shipped fully assembled in a metal enclosure. Your package
should contain the following items:
• Stationary stroboscope
• Flash tube, factory installed
• Operating instruction
• Connector (1) for trigger input and supply
After unpacking, please examine your stroboscope for signs of mishandling during
shipping. If damage has occurred, contact Electromatic before attempting to plug in the
instrument. Notify your shipping carrier immediately for damage claim instructions.
2.0 Safety
LS-9-12000 stroboscopes operate at dangerously high voltages. Disconnect
the unit from the power supply and wait two minutes for the capacitor to
discharge before replacing the lash tube. Any service beyond lash tube
replacement is not recommended.
•
Do not look directly at the emitted light; it can damage your eyes. It
may also trigger seizures in people with photosensitive epilepsy.
•
Flash tubes are illed with high-pressure gas. When handling the
lash tube always always wear safety glasses and protective gloves.
•
Do not allow inlammable liquids and water enter the stroboscope.
•
When operating the stroboscope with the front screen facing
upwards, the screen can heat up.
•
The stroboscope must not be used in explosion hazard areas.
– 2 –
– 7 –
6.0 Flash Tube Replacement
1. Switch off the LS-9-12000 and remove the power supply plug.
Wait two minutes for the capacitor to discharge before continuing.
2. Loosen the six (6) T10-TORX screws on the side of the unit with the
yellow label.
NOTE: Do not open the
screws on the other side of the
unit. They are not T10-TORX
type.
3. Slide front glass out laterally.
4. Using a clean cloth, carefully
remove the old lashbulb.
5. Pick up new lashbulb with a
clean cloth (free of dust and
grease) and insert into the two
retaining clips. Orient the tube
so that the trigger wire can
be easily attached to the
terminal clamp screw.
Ensure that the lashtube is
correctly positioned inthe socket.
IMPORTANT: When the lashtube is
correctly positioned, the extra length
of wire will be closest to the RIGHT
retaining clip and near the terminal
clamp screw.
6. Re-insert the front screen and replace the screws.
NOTE: Never operate the stroboscope without a front screen or with a
defective front screen.
High Voltage!
Use only the manufacturer’s original lashtubes.
Never touch the lashtube to your bare skin.
Only use the device with an undamaged front lens itted.