Glass jar installation – Cannon Instrument CT-1000 (Contant Temperature Bath) User Manual
Page 8
6
CANNON
®
CT-1000 Constant Temperature Bath Instruction & Operation Manual
Revision 3.1e—June, 2011;
CANNON
®
Instrument Company
2139 High Tech Road • State College, PA • USA 16803
2. Move the bath
unit housing to its
permanent loca-
tion on the
laboratory bench.
3. Remove all eight
screws from the
stainless steel top
covers (see Figure
3).
4. Disconnect all
probes, heaters,
and float switch
from the rear
panel.
5. Remove the front top cover and rear top cover (see Figure 4 and 5).
Use caution when removing the rear top cover because the tempera-
ture control probes and heating elements are attached to it.
Glass jar installation
1. Remove the glass jar from its
box.
WARNING
The glass jar is heavy. Use
caution when lifting it.
2. Make sure the rubber support
ring is seated properly around
the bottom jar opening in the
bath unit (see Figure 6).
Figure 3: Removing screws from top cover
Figure 4:
Removing front top cover
Figure 5:
Removing rear top cover
Figure 6: Checking support ring