6 Operation

24

K-355 Operation Manual, Version G

6 .5

Steam generator status

6 .5 .1

Heating

A constant colon (:) and an “H” in the STEAM-display indicates, that the steam generator is heating.
A flashing colon (:) and a signal sound for 1 second indicates that the steam generator is ready.

6 .5 .2

Standby mode (after 30 minutes)

The TIME display and STEAM display show (-----).
Press “START” to cancel the standby mode.

6 .6

Switching ON/OFF checklist

To switch on and preheat the instrument, proceed as follows:

Check the filling level of the supply tanks.

Open the main water tap.

Switch on the instrument and wait until the steam generator is ready.

Insert an empty sample tube and an empty receiving vessel.

Close the protective door.

Set the distillation time to 2 minutes (see section “Setting the distillation time”) and the steam to

100% (see section “Setting the steam power”).
Start the distillation by pressing “START”.

To switch off the instrument, proceed as follows:

Switch off the instrument.

Close the main water tap.

Rinse the reagent 1 pump with distilled water when it is used for the additon of strong acids.

6 .7

Carrying out a distillation

To carry out a distillation, proceed as follows:

Switch on the instrument and wait until the steam generator is ready.

Insert the sample tube containing the sample.

Prepare a conical flask or a similar receptacle with a volume of about 250 ml as receiving vessel.

If required, select suitable chemicals in accordance with the method and the substance you deter-

mine.
Add reagents as required.

Set the required distillation time or select a program.

Start the distillation by pressing “START”.


For Kjeldahl distillation note the following:

Fill boric acid solution of 4% and 2-3 drops of Sher indicator into the receiving vessel. Alternative:

Select suitable chemicals for the back titration.
Add water to dilute the sample by pressing “REAGENT 1”. The dilution with water prevents a too

violent reaction when the sodium hydroxide solution is subsequently added. The sample must be
diluted in a ratio of approximately 1 part of acid to 2 parts of water.
Add sodium hydroxide solution by pressing “REAGENT 2”: Recommended alkali concentration:

32%. Add alkali to the sample in a ratio of about 1 part of acid to 3 parts of sodium hydroxide