Power Plate pro7 User Manual
Page 8

Guidance and manufacture’s declaration – electromagnetic immunity –
for ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacture’s declaration – electromagnetic immunity
The Pro7 is intended for use in the electromagnetic environment specified below. The customer or the user of
Pro7 should assure that it is used in such an environment.
Immunity test
IEC 60601 test
Compliance
Electromagnetic environment
level
level -
guidance
Portable and mobile RF communications
equipment should be used no closer to
any part of the Pro7, including cables, than
the recommended separation distance
calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
Conducted RF
3 Vrms
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
3 Vrms
80 MHz to 800 MHz
IEC 61000-4-3
80 MHz to 2.5 GHz
800 MHz to 2.5 GHz
Where
P is the maximum output power rating
of the transmitter in watts (W) according
to the transmitter manufacturer and d is the
recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
a
should be less than the compliance level in each
frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the Pro7 is used exceeds
the applicable RF compliance level above, the Pro7 should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as reorienting or relocating the Pro7.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
P
d
167
.
1
=
P
d
167
.
1
=
P
d
167
.
1
=
!
Electromagnetic Compatibility (EMC)
Precautions
The pro7™ model is suitable for use in
all establishments, including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.
Both models have been tested against,
and have passed the applicable
requirements of relevant electromedical
standards, including EN 60601-1-2:2007.
The pro7 model emits electromagnetic
energy to perform its intended function.
Nearby electronic equipment may be
affected by this emission. Similarly
the pro7 model may be affected by
electromagnetic emissions from other
equipment in the vicinity.
The pro7 model is intended for use in the
electromagnetic environment specified
below. The customer or user of the pro7
model should ensure that it is used in
such an environment:
– Floors should be wood, concrete or
ceramic tile. If the floor is covered with
synthetic material, the relative humidity in
the room should be at least 30%.
– Power quality should be that of a typical
domestic, commercial or hospital
environment as appropriate.
– The pro7 model should not be used
during power interruptions.
– Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
The pro7 model is intended for use in an
electronic environment in which radiated
radio-frequency (RF) disturbances
are controlled.
The customer or user of the pro7 model
can help prevent electromagnetic
interference by maintaining a minimum
distance between portable and mobile
RF communications equipment
(transmitters) and the pro7 model as
recommended in this user manual,
according to the maximum output power
of the communications equipment:
The adequate separation distance
to provide base immunity to RF
disturbances is 1.0m.
•
USB Port - USB port supplies 5Volt DC
current. This port is only used for soft-
ware updates and uploading personal
pictures to user profiles. The operator
must take care to not touch the USB port
and a patient simultaneously.
•
LAN port - the pro7 has included a
LAN port on the lower back side of
the machine. This LAN port is used to
connect a LAN cable to access the
internet.The operator must take care to
not touch the LAN port and a patient
simultaneously.
•
Phone jack adapter- On the Control
Panel of the pro7, there is a phone
jack adapter for sound playback. The
operator must take care to not touch
the Phone jack adapter and a patient
simultaneously.
Retain this user manual for your future
reference.