Guidance and manufacture’s declaration – electromagnetic emissions-

for all ME EQUIPMENT and ME SYSTEMS

Guidance and manufacture’s declaration – electromagnetic emission

The Pro7 is intended for use in the electromagnetic environment specified below. The customer or the user
of the Pro7 should assure that it is used in such an environment.

Emission test

Compliance

Electromagnetic environment – guidance

Group 1

The Pro7 use RF energy only for its internal function.

Therefore, its RF emissions are very low and are not

likely to cause any interfe rence in nearby electronic

equipment.

RF emission CISPR 11

Class B

Class A with IEC61000-3-2

Complies with IEC61000-3-3

Harmonic emissions

Class A

The Pro7 is suitable for use in all establishments,

IEC 61000-3-2

including domestic establishments and those directly

connected to the public low-voltage power supply

Voltage fluctuations/

Comply

network that supplies buildings used for domestic

flicker emissions

purposes.

IEC 61000-3-3


Wi-Fi Specifications

Network Standard: IEEE 802.11b/g/n

Band: 2.400-2.4835GHz

Cover Range: 20meter

Wi-Fi Power: 10dBm

EMC Declaration

Guidance and manufacture’s declaration – electromagnetic immunity –

for all ME EQUIPMENT and ME SYSTEMS

Guidance and manufacture’s declaration – electromagnetic immunity

The Pro7 is intended for use in the electromagnetic environment specified below. The customer or the user of
Pro7 should assure that it is used in such an environment.

Immunity test

IEC 60601 test level

Compliance level

Electromagnetic environment - guidance

Electrostatic

±6 kV contact

±6 kV contact

Floors should be wood, concrete or

discharge (ESD)

±8 kV air

±8 kV air

ceramic tile. If floor are covered with

IEC 61000-4-2

synthetic material, the relative humidity

should be at least 30%.

Electrical fast

±2 kV for power

±2kV for power

Mains power quality should be that of

transient/burst

supply lines

supply lines

a typical commercial or hospital IEC

61000-4-4

environment.

Surge

± 1 kV line(s) t

±1 kV differential

Mains power quality should be that of

IEC 61000-4-5

o line(s)

mode

a typical commercial or hospital

environment.

Voltage dips, short

<5% U

T

<5% U

T

Mains power quality should be that

interruptions and

(>95% dip in U

T

)

(>95% dip in U

T

)

of a typical commercial or hospital

voltage variations

for 0.5 cycle

for 0.5 cycle

environment. If the user of the Pro7

on power supply

requires continued operation during

input lines

40% U

T

40% U

T

power mains interruptions, it is IEC

61000-4-11

(60% dip in U

T

)

(60% dip in U

T

)

recommended that the Pro7 be powered

for 5 cycles

for 5 cycles

from an uninterruptible power supply or a

battery.

70% U

T

70% U

T

(30% dip in U

T

)

(30% dip in U

T

)

for 25 cycles

for 25 cycles

<5% U

T

<5% U

T

(>95% dip in U

T

)

(>95% dip in U

T

)

for 5 sec

for 5 sec

Power frequency

3A/m

3A/m

Power frequency magnetic fields should

(50Hz) magnetic

be at levels characteristic of a typical

61000-4-8

location in a typical commercial or

hospital environment.

NOTE U

T

is the a.c. mains voltage prior to application of the test level.