Electromagnetic Compatibility (EMC)

Precautions

The pro6

+

model is suitable for use in

all establishments, including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.
Both models have been tested against,
and have passed the applicable
requirements of relevant electromedical
standards, including EN 60601-1-2:2007.

The pro6

+

model emits electromagnetic

energy to perform its intended function.
Nearby electronic equipment may be
affected by this emission. Similarly
the pro6

+

model may be affected by

electromagnetic emissions from other
equipment in the vicinity.

The pro6

+

model is intended for use

in the electromagnetic environment
specified below. The customer or user of
the pro6

+

model should ensure that it is

used in such an environment:

– Floors should be wood, concrete or

ceramic tile. If the floor is covered with
synthetic material, the relative humidity in
the room should be at least 30%.

– Power quality should be that of a typical

domestic, commercial or hospital
environment as appropriate.

– The pro6

+

model should not be used

during power interruptions.

– Power frequency magnetic fields should

be at levels characteristic of a typical
location in a typical commercial or
hospital environment.

The pro6

+

model is intended for use in an

electronic environment in which radiated
radio-frequency (RF) disturbances

are controlled.

The customer or user of the pro6

+

model can help prevent electromagnetic
interference by maintaining a minimum
distance between portable and mobile
RF communications equipment
(transmitters) and the pro6

+

model as

recommended in this user manual,
according to the maximum output power
of the communications equipment:

The adequate separation distance
to provide base immunity to RF
disturbances is 1.0m.

USB Port - USB port supplies 5Volt DC
current. This port is only used for soft-
ware updates and uploading personal
pictures to user profiles. The operator
must take care to not touch the USB port
and a patient simultaneously.

LAN port - the pro6

+

has included a

LAN port on the lower back side of
the machine. This LAN port is used to
connect a LAN cable to access the
internet.The operator must take care to
not touch the LAN port and a patient
simultaneously.

Phone jack adapter- On the control
panel of the pro6

+

there is a phone

jack adapter for sound playback. The
operator must take care to not touch
the phone jack adapter and a patient
simultaneously.

Retain this user manual for your future
reference.