Power Plate pro6 plus User Manual
Page 13

Guidance and manufacture’s declaration – electromagnetic immunity –
for all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture’s declaration – electromagnetic immunity
The pro6
+
is intended for use in the electromagnetic environment specified below. The customer or user of the
pro6
+
should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment - guidance
Electrostatic
±6 kV contact
±6 kV contact
Floors should be wood, concrete or
discharge (ESD)
±8 kV air
±8 kV air
ceramic tile. If floor are covered with
IEC 61000-4-2
synthetic material, the relative humidity
should be at least 30%.
Electrical fast
±2 kV for power
±2kV for power
Main power quality should be that of
transient/burst
supply lines
supply lines
a typical commercial or hospital IEC
61000-4-4
environment.
Surge
± 1 kV line(s) t
±1 kV differential
Main power quality should be that of
IEC 61000-4-5
o line(s)
mode
a typical commercial or hospital
environment.
Voltage dips, short
<5% U
T
<5% U
T
Main power quality should be that
interruptions and
(>95% dip in U
T
)
(>95% dip in U
T
)
of a typical commercial or hospital
voltage variations
for 0.5 cycle
for 0.5 cycle
environment. If the user of the pro6
+
on power supply
requires continued operation during
input lines
40% U
T
40% U
T
power main interruptions, it is IEC 61000-
4-11
(60% dip in U
T
)
(60% dip in U
T
)
recommended the pro6
+
be powered
for 5 cycles
for 5 cycles
from an uninterruptible power supply or a
battery.
70% U
T
70% U
T
(30% dip in U
T
)
(30% dip in U
T
)
for 25 cycles
for 25 cycles
<5% U
T
<5% U
T
(>95% dip in U
T
)
(>95% dip in U
T
)
for 5 sec
for 5 sec
Power frequency
3A/m
3A/m
Power frequency magnetic fields should
(50Hz) magnetic
be at levels characteristic of a typical
61000-4-8
location in a typical commercial or
hospital environment.
NOTE U
T
is the a.c. main voltage prior to application of the test level.
11 of 22