Power Plate my5 User Manual
Page 8

the applicable requirements of relevant
electromedical standards, including EN
60601-1-2:2007.
The my5™ model must emit
electromagnetic energy in order to
perform its intended function. Nearby
electronic equipment may be affected
by this emission. Similarly the my5™
model may itself be affected by
electromagnetic emissions from other
equipment in the vicinity.
The my5™ model is intended for use
in the electromagnetic environment
specified below. The customer or user of
the my5™ model should assure that it is
used in such an environment:
– Floors should be wood, concrete or
ceramic tile. If the floor is covered with
synthetic material, the relative humidity
in the room should be at least 30%.
– Mains power quality should be that of a
typical domestic, commercial or hospital
environment as appropriate.
– The my5™ model should not be used
during power mains interruptions.
– Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.
The my5™ model is intended for
use in an electronic environment in
which radiated radio-frequency (RF)
disturbances are controlled.
The customer or user of the my5™
model can help prevent electromagnetic
interference by maintaining a minimum
distance between portable and mobile RF
communications equipment (transmitters)
and the my5™ model as recommended
below, according to the maximum output
power of the communications equipment:
The adequate separation distance
to provide base immunity to RF
disturbances is 1.0 metres.
Retain this user manual for your future
reference.
•
•
•
•