AirControl C LED

Part Number

819703

819704

1 green LED for normal air speed

1 green LED for normal air speed

Display – Visual

1 red LED flashing for alarm

1 red LED flashing for alarm

No digital display

3 digit display with velocity reading

Units

N/A

meter per second (m/s) / FPM

Display Range

N/A

0 - 2.00 m/s / 0 - 400 FPM

Alarm Setpoint

Standard: below 0.39m/s / 78 FPM

Standard: below 0.39m/s / 78 FPM

Alarm Delay

Selectable: 15s or 30s

Selectable: 15s or 30s

Analog Output

0-10V 0-10V

Alarm Indication

1 red LED flashing & audible buzzer

1 red LED flashing & audible buzzer

Alarm Mute

Light On/Off

Night Setback On/Off

Alarm Relay

Yes, optional

Yes, optional

Battery Back Up

Yes, optional

Yes, optional

Sash High Input

Audible and orange flashing LED indicate Audible and orange flashing LED indicate

sash position switch has been tripped

sash position switch has been tripped

Mounting

Flush or surface box (option)

Flush or surface box (option)

Calibration

Factory pre-calibrated @ 0.5m/s / 100 FPM Factory pre-calibrated @ 0.5m/s / 100 FPM

Recalibration possible

Recalibration possible

Power Requirement

12 VDC (power supply included)

12 VDC (power supply included)

Orientation

Vertical / Horizontal Vertical

Front fascia: 210L x 90W x 10D mm

Front fascia: 210L x 90W x 10D mm

Monitor Dimensions (metric)

Surface box: 205L x 85W x 14D mm

Surface box: 205L x 85W x 14D mm

Front fascia: 8.27"L x 3.54"W x 0.39"D Front fascia: 8.27"L x 3.54"W x 0.39"D

Monitor Dimensions (U.S.)

Surface box: 8.07"L x 3.35"W x 0.55"D Surface box: 8.07"L x 3.35"W x 0.55"D