Type junior, Laboratory airflow monitor, Contacts and wiring – Plastec Junior Air Control User Manual
Page 9
Laboratory Airflow Monitor
TYPE JUNIOR
Contacts and Wiring
2216 60th Drive East • Bradenton, Florida 34203
(941) 751-7596 • Fax (941) 751-7598 •
www.plastecvent.net
9
MAINTENANCE
Front plate and stickers of airflow monitor may be cleaned with mild soap and water on a damp
cloth to remove finger marks, oils and residue. Do not use abrasives. Do not allow liquids to enter
the plastic casing. Dry the monitor thoroughly after cleaning.
TROUBLESHOOTING
PROBLEM
CHECK
No indicators
Power supply may not be plugged into AC supply.
Improper alarm set point
Airflow sensor is factory precalibrated at 0.5 m/s or 100 FPM. Field
reading may be different if sensor is not properly located or hood
design affects stable airflow. If intended air speed is different than
0.5 m/s / 100 FPM or if actual anemometer reading is different,
follow Sensor Recalibration Procedure page 1.