beautypg.com

Type junior, Laboratory airflow monitor, Specifications – Plastec Junior Air Control User Manual

Page 5

background image

5

Junior AirControl

Part Number

819699

1 green LED for normal air speed

Display – Visual

1 red LED flashing for alarm

No digital display

Alarm Setpoint

Standard: below 0.39m/s / 78 FPM

Alarm Indication

1 red LED flashing and audible buzzer

Alarm Mute

Mounting

Flush or surface box

Calibration

Factory pre-calibrated @ 0.5m/s / 100 FPM

Recalibration possible

Power Requirement

12 VDC (power supply included)

Orientation

Vertical only

Front fascia: 100L x 70W x 10D mm

Monitor Dimensions (metric)

Surface box: 100L x 20W x 20D mm

Front fascia: 3.94" x 2.76" x 0.39"

Monitor Dimensions (U.S.)

Surface box: 3.94" x 0.79" x 0.79"

2216 60th Drive East • Bradenton, Florida 34203

(941) 751-7596 • Fax (941) 751-7598 •

www.plastecvent.net

5

Laboratory Airflow Monitor

TYPE JUNIOR

Specifications