Troubleshooting specifications, Problems solutions, System specifications – Hudson Medical Alternating Pressure Pad User Manual
Page 6: Pump - #8202, Bubble pad - #8201

Problems
Solutions
The pump
doesn’t work
Incomplete
inflation
(low pressure)
1. Check if the plug is inserted firmly into the outlet.
2. Turn on the power switch again.
• If the power indicator is ON and the pump
still doesn’t work, contact your local dealer
immediately.
• If the power indicator is OFF, there may be
a faulty outlet. Try to connect the power
cord to another outlet. If the power indicator
is still OFF, contact a qualified electrician for
main power check.
1. Adjust the pressure knob to Max.
2. Check to see if the inflation tubes connected to
the pump are twisted or disconnected, then
check for any air leakage.
• Always keep the tubes straight.
• Change the tubes if there is any leakage.
Troubleshooting
Specifications
System
Specification
Classification
Applied Part
Power Supply*
Environment
Requirements
Class II; AP/APG NO; IPX0; Type B
Bubble Pad
AC 120V 60Hz; 12W
• Temperature: Operation 50°F~95°F (10~35°C)
Storage 5°F ~122°F (-15~50°C)
Shipping 5°F ~158°F (-15~70°C)
• Humidity: Operation 20%~80% non-condensing
Storage 10%~90% non-condensing
System Specifications
* AP/APG NO indicates the device is NOT suitable for use in the presence of
flammable anesthetic mixture with air or with oxygen or nitrous oxide.
** Type B symbol indicates the degree of protection against electric shock.
*** Above specifications are applicable to those areas operating per same power
supply range.
Pump
Specification
Dimensions
Weight
Pressure Range
Cycle Time
10.0 x 5.1 x 3.75 in
3.5 lbs
30-60mmHg
6 min/60Hz
Pump - #8202
8201 Bubble pad w/ extended bed flaps 78 x 34 x 2.5 PVC 300 lbs.
Bubble Pad - #8201
Part Arrangement Dimension Material Weight
No.
Support