Troubleshooting specifications, Problems solutions, System specifications – Hudson Medical Alternating Pressure Pad User Manual

Page 6: Pump - #8202, Bubble pad - #8201

background image

Problems

Solutions

The pump

doesn’t work

Incomplete

inflation

(low pressure)

1. Check if the plug is inserted firmly into the outlet.

2. Turn on the power switch again.
• If the power indicator is ON and the pump

still doesn’t work, contact your local dealer

immediately.

• If the power indicator is OFF, there may be

a faulty outlet. Try to connect the power

cord to another outlet. If the power indicator

is still OFF, contact a qualified electrician for

main power check.

1. Adjust the pressure knob to Max.

2. Check to see if the inflation tubes connected to

the pump are twisted or disconnected, then

check for any air leakage.

• Always keep the tubes straight.

• Change the tubes if there is any leakage.

Troubleshooting

Specifications

System

Specification

Classification
Applied Part
Power Supply*
Environment

Requirements

Class II; AP/APG NO; IPX0; Type B
Bubble Pad
AC 120V 60Hz; 12W
• Temperature: Operation 50°F~95°F (10~35°C)

Storage 5°F ~122°F (-15~50°C)

Shipping 5°F ~158°F (-15~70°C)

• Humidity: Operation 20%~80% non-condensing

Storage 10%~90% non-condensing

System Specifications

* AP/APG NO indicates the device is NOT suitable for use in the presence of

flammable anesthetic mixture with air or with oxygen or nitrous oxide.

** Type B symbol indicates the degree of protection against electric shock.

*** Above specifications are applicable to those areas operating per same power

supply range.

Pump

Specification

Dimensions
Weight
Pressure Range
Cycle Time

10.0 x 5.1 x 3.75 in
3.5 lbs
30-60mmHg
6 min/60Hz

Pump - #8202

8201 Bubble pad w/ extended bed flaps 78 x 34 x 2.5 PVC 300 lbs.

Bubble Pad - #8201

Part Arrangement Dimension Material Weight

No.

Support