Ground contact, Bnc connector – ETS-Lindgren 3625-2 LISN User Manual

Page 14

background image

14

Installation and Application

www.ets-lindgren.com

Ground Contact

An RF ground stud is provided on the front panel and back panel for

bonding to the ground plane.

Ground contact can be made through the receptacle marked GND or through the

brass stud on the back panel.

If a ground contact to the Equipment Under Test (EUT) is made to the power

supply ground, it may be made through the GND receptacle or the ground stud

on the front panel. No ratio isolation is provided in this line.

Load side of the Model 3625/2 should be made through the two-wedge type

receptacles on the front panel.

BNC Connector

The BNC coaxial connector of the line under test should be connected to the field

intensity meter or spectrum analyzer input. All other BNC receptacles must be

terminated with a 50-ohm RF termination.

Removal of a terminator and connection of the BNC receptacles will not generate

power surges.