1.6 Troubleshooting

Symptom

Possible Cause

Remedy

Cannot turn on

Batteries are low
Incorrectly installed battery

Replace batteries, check
polarity

Poor weighing accuracy Improper calibration

Unstable environment

Scale was not zeroed prior to
use

Tare function is active

Displaying incorrect weight

Perform calibration

Move scale to stable
location
Press the ZERO key to
zero the scale prior to
use

Press Tare key to
remove tare weight held
in memory

Ensure correct unit of
measure is active

CAL. Er displayed

Unstable environment

Incorrect calibration weight
used

Move scale to suitable
location

Use correct calibration
weight

Err2 displayed

Overload Zero Error

Normal Calibration condition

Load cell is damaged

Excess weight applied to
scale upon zero or power
on, remove weight

Remove weight

Err1 displayed

Under load Zero Error

Verify weight platter is
installed correctly, power
off then on

Perform Calibration

_ _ _ _ _ displayed

Overload Condition

Excess weight applied to
scale, remove weight