1.6 Troubleshooting
Symptom
Possible Cause
Remedy
Cannot turn on
Batteries are low
Incorrectly installed battery
Replace batteries, check
polarity
Poor weighing accuracy Improper calibration
Unstable environment
Scale was not zeroed prior to
use
Tare function is active
Displaying incorrect weight
Perform calibration
Move scale to stable
location
Press the ZERO key to
zero the scale prior to
use
Press Tare key to
remove tare weight held
in memory
Ensure correct unit of
measure is active
CAL. Er displayed
Unstable environment
Incorrect calibration weight
used
Move scale to suitable
location
Use correct calibration
weight
Err2 displayed
Overload Zero Error
Normal Calibration condition
Load cell is damaged
Excess weight applied to
scale upon zero or power
on, remove weight
Remove weight
Err1 displayed
Under load Zero Error
Verify weight platter is
installed correctly, power
off then on
Perform Calibration
_ _ _ _ _ displayed
Overload Condition
Excess weight applied to
scale, remove weight