beautypg.com

Iris Sample Processing StatSpin® CenSlide 2000 Urine Sediment Analysis System User Manual

Page 7

background image

Product Overview

OPERATOR’S MANUAL

CenSlide 2000 Urinalysis System

Page 1 - 4

1.3 CenSlide 2000 Centrifuge Specifications

Power Requirements

Model

Power Supply Mains

M401-12 . . . . . . . . . . . . . . .120 VAC, 60 Hz, 26 W

(Product No. 2C-9000-001)

M401-22 . . . . . . . . . . . . . . .90-240 VAC, 50/60 Hz, 26 W

(Product No. 2C-9000-22)

Centrifuge Motor . . . . . . . .24 VDC, Brush Type,

Solid State Microprocessor
Controlled

Lid and Bowl . . . . . . . . . . .One piece molded Nylon plastic

Centrifuge Cycle
Ramp to 1350 rpm ±10% . .

15 seconds

Hold 1350 rpm . . . . . . . . . .

15 seconds

Ramp down to 400 rpm . . .

15 seconds

Flick tube . . . . . . . . . . . . . .

2 seconds

Centrifugal Force . . . . . . .97 xg

Power Consumption . . . . .20 W

Physical Dimensions

Height: . . . . . . . . . . . . . . . .5.6"

(14.2 cm)

Width: . . . . . . . . . . . . . . . . .11.4" (29.0 cm)
Depth: . . . . . . . . . . . . . . . . .12.9" (32.8 cm)
Weight: . . . . . . . . . . . . . . . .9 lbs. (4.1 Kg)

Environmental Conditions

Indoor Use • Altitude up to 2000m • Temperature 5°C to 40°C

Maximum relative humidity 80% for temperatures up to 31°C
decresasing linearly to 50% relative humidity at 40°C

Main supply voltage fluctuations not to exceed ± 10%
of the nominal voltage

Transient overvoltages according to installation category II

Pollution degree 2

Rotor . . . . . . . . . . . . . . . . .6 Position, fixed 45° angle,

Nylon plastic

Switches . . . . . . . . . . . . . .Pressure sensitive membrane