1.3 CenSlide 2000 Centrifuge Specifications
Power Requirements
Model
Power Supply Mains
M401-12 . . . . . . . . . . . . . . .120 VAC, 60 Hz, 26 W
(Product No. 2C-9000-001)
M401-22 . . . . . . . . . . . . . . .90-240 VAC, 50/60 Hz, 26 W
(Product No. 2C-9000-22)
Centrifuge Motor . . . . . . . .24 VDC, Brush Type,
Solid State Microprocessor
Controlled
Lid and Bowl . . . . . . . . . . .One piece molded Nylon plastic
Centrifuge Cycle
Ramp to 1350 rpm ±10% . .
≈
15 seconds
Hold 1350 rpm . . . . . . . . . .
≈
15 seconds
Ramp down to 400 rpm . . .
≈
15 seconds
Flick tube . . . . . . . . . . . . . .
≈
2 seconds
Centrifugal Force . . . . . . .97 xg
Power Consumption . . . . .20 W
Physical Dimensions
Height: . . . . . . . . . . . . . . . .5.6"
(14.2 cm)
Width: . . . . . . . . . . . . . . . . .11.4" (29.0 cm)
Depth: . . . . . . . . . . . . . . . . .12.9" (32.8 cm)
Weight: . . . . . . . . . . . . . . . .9 lbs. (4.1 Kg)
Environmental Conditions
Indoor Use • Altitude up to 2000m • Temperature 5°C to 40°C
Maximum relative humidity 80% for temperatures up to 31°C
decresasing linearly to 50% relative humidity at 40°C
Main supply voltage fluctuations not to exceed ± 10%
of the nominal voltage
Transient overvoltages according to installation category II
Pollution degree 2
Rotor . . . . . . . . . . . . . . . . .6 Position, fixed 45° angle,
Nylon plastic
Switches . . . . . . . . . . . . . .Pressure sensitive membrane