Iris Sample Processing StatSpin® CenSlide 2000 Urine Sediment Analysis System User Manual
Page 27

Trouble Shooting Guide
OPERATOR’S MANUAL
CenSlide 2000 Urinalysis System
Page 5 - 3
5.0 Trouble Shooting
OBSERVATIONS
ACTIONS
The lid to the centrifuge is open
Disconnect the power cord at the
but the green light is not flashing.
back of the unit and then reconnect it.
This will cause a power "RESET".
The yellow light will not go on
Check that there is power to the
when the start button is pressed.
centrifuge by observing the green light.
Check that the lid is securely latched.
Microscopic observation
Make sure the CenSlide tubes have at least
and focus difficulty.
a one minute settling time in the horizontal
position after centrifugation.
Assure that the CenSlide tubes are properly
placed in the CenSlide Holder.
Using the 10x Objective, focus the
microscope on the top or bottom of the
CenSlide tube surface and then slowly
adjust up or down until you find the
observation plane. Then switch to your 40x
objective. If it comes in contact with the
CenSlide tube, you will need to obtain a
long working distance objective through
your local distributor or directly through
StatSpin.
If you have any questions, or require customer assistance,
please call 1-800-782-8774 from 8:00 am to 6:00 pm EST.