System overview – Iris Sample Processing StatSpin® CytoFuge 2 Compact Cytocentrifuge System User Manual
Page 6

2
Section
2
System Overview
Principle and Intended Use
For in vitro diagnostic use in concentrating cells onto a microscope slide.
The CytoFuge 2 is a microprocessor-controlled cell preparation system that uses centrifugal force to
deposit cells onto microscope slides. Samples are centrifuged in reusable or disposable gasket-
sealed chambers (Cell Concentrators), or in disposable chambers (Filter Concentrators) that include
a filter material to absorb and capture suspension fluid during cytocentrifugation
A four-position rotor with transparent cover is supplied as standard equipment. The system is
designed to be operated on a laboratory bench or, if potentially biohazardous material is processed,
in a biological safety cabinet. A cover interlock system prevents operation until the centrifuge cover
is securely latched, and prevents access to spinning parts until the rotor has come to a complete
stop. Speeds from 600 rpm to 4400 rpm and cycle times from 2 minutes to 30 minutes may be
selected. Automatic acceleration rate control and dynamic braking are included to protect delicate
samples.
Symbols and Definitions
Start button
The start button initiates a pre-timed cycle at the
speed selected. NOTE: The CytoFuge 2 does not
have an on/off switch and is normally left plugged in
and “on”.
Stop/open button
The stop/open button interrupts the cycle and stops
the centrifugation. This button may also be used to
release the cover.
•
Power indicator
The green LED identified as power is illuminated
when the CytoFuge 2 is connected to a power source.
Release cover
Located below stop/open button on instrument front
panel
Time selector
Available time settings: 2, 4, 6, 8, 10, and 30 minutes
Speed selector
Available speed settings: 600, 700, 850, 1000, 1300,
1600, 2200, and 4400 rpm
Power input
DC Power Input: 24V DC Plug-in
IVD