Maintenance – Iris Sample Processing StatSpin® CytoFuge 2 Compact Cytocentrifuge System User Manual
Page 12

8
Section
5
Maintenance
Overview
StatSpin recommends that instrument operators perform periodic inspections and preventative
maintenance on all StatSpin instruments. Contact StatSpin’s customer service department or
distributor if, at any time, the instrument is not functioning properly.
CAUTION - Unplug the CytoFuge 2 from the wall outlet before performing maintenance.
WARNING - Do not expose CytoFuge 2 and its rotor to strong or concentrated acids,
bases, esters, aromatic or halogenated hydrocarbons, ketones or strong
oxidizing agents.
Cleaning
Clean the outside surfaces and switch overlay panel with a water-dampened cloth and mild
detergent. Clean the inner surface or bowl with a mild detergent, and if necessary, a
disinfectant, wiping surfaces with a
dampened
cloth using 70% alcohol or 10% bleach
solution.
CAUTION - DO NOT spray the bowl or outer surfaces with detergent or bleach. Excess
liquid will harm the electronics and subsequent problems will not be covered under
warranty.
Checking the Rotor Speed
The rated speeds (
+
5%) can be checked with a photoelectric tachometer available from many
sources. If the CytoFuge 2 fails to achieve operating speed
+
5% contact your distributor or StatSpin
Customer Service department.
Service
There are no user-serviceable parts. Refer all service to qualified service personnel.
Reference the StatSpin Warranty for further instruction. Be sure to complete and return the warranty
card as directed.
Decontamination before returning for service
Any instrument or accessory containing accumulated blood and/or other biological or chemical
deposits must be cleaned prior to shipment to the manufacturer/dealer for service. This
decontamination is required by Federal Law (Title 48 and 49 of the Federal Regulations) and in
accordance with the Environmental Protection Agency’s Regulations for Biohazard Waste
Management. StatSpin personnel cannot perform this decontamination.