beautypg.com

Drucker Diagnostics QBC Horizon 755VES Centrifuge (120V, 60Hz) User Manual

Page 24

background image

Section 8

CALIBRATION CHECKS

According to Federal Regulations, centrifuges, such as the QBC Capillary Centrifuge, require verification or
calibration as follows: 1) before initial use; 2) after repair or adjustments; and 3) annually after use. The centrifuge
timer, however, should be checked for accuracy at least every 3 months.

The QBC Capillary Centrifuge has been fully

calibrated at the factory and is ready to deliver consistent, accurate

test results upon first use. For subsequent calibration checks, follow the procedures in 8.1 and 8.2 below.

8.1 CHECKING ROTOR SPEED

The rotor of the QBC Capillary Centrifuge is designed to spin at 12,000 ± 80 rpm.

At least every 6 months or after

any adjustments or repairs, check the rotor speed with a non-contact strobe tachometer to assure that speed
is sufficient to produce visibly distinct and differentiated cell layers in the QBC tube. A window in the top lid and a
reference line on the knob of the rotor cover are provided for speed checks with a non-contact tachometer.

8.2 CHECKING TIMER ACCURACY

The electronic timer of the Centrifuge is designed to be accurate to 300 ± 15 s. Check the timer for accuracy
against a reliable stopwatch or quartz timer at least every

3 months as follows:

a) Install the rotor cover.
b) Close the latch and top lid.
c) Press and release the ON button, and

simultaneously start the stopwatch.

d) Stop the

stopwatch when you audibly detect cutoff of power to the motor and the centrifuge begins to

decelerate.

e) The stopwatch reading should be 5 min ± 15 s.

If the 15-s tolerance is exceeded, repeat the check to eliminate the possibility of procedural errors. Request
service if the accuracy check is unsatisfactory.

19

This manual is related to the following products: