Dako Hybridizer User Manual
Page 4

Corporate Headquarters
Denmark
Tel +45 44 85 95 00
Distributors in more
than 50 countries
www.dako.com
00187
25APR07
Run a Denaturation and Hybridization Program (Denat & Hyb)
1. Select Run a PGM
2. Select a PGM
n
Cursor highlights “Run a
PGM” line
n
Press “Enter” button to accept
n
Scroll through program
Nos. 1-40/program names
n
Press “Enter” button to accept
n
Display will confirm PGM No./
name and Denat & Hyb times
and temperatures
n
Cursor highlights “Run PGM”
line
n
Press “Enter” button to accept
3. Insert Humidity Control Strips and Add Slides
4. Heating to Denat Temp
n
Insert two Humidity Control
Strips into the inside slide lid
n
Saturate strips with distilled
water or equivalent
n
Add slides and close lid
n
Cursor highlights “Start” line
n
Press “Enter” button to run the
program
n
Display indicates present
temperature of the slides
5. Denaturation
6. Cooling to Hyb Temp
n
The Hybridizer will beep twice
once temperature reaches
denaturation set point
n
Denaturation time will count
down from the set time
n
The Hybridizer will
automatically cool to
hybridization set temperature
once denaturation is
completed
7. Hybridization
8. Process Completed
n
Hybridization time will count
down from the set time
once temperature reaches
hybridization set point
n
The Hybridizer will beep
five times upon program
completion
n
Cursor highlights “End PGM/
Main Menu”
n
Press “Enter” to end program
n
See Handbook for further
details and for running a
“Hyb Only” or “Fixed Temp”
program
Run a PGM
Edit a PGM
Create a PGM
Present Temp: 37 °C
PGM 01 HER2
82 °C :05; 45 °C 20:00
Run PGM
Main Menu
PGM 01 HER2
Add Slides-Close Lid
Start
Main Menu
PGM 01 HER2
**Heating**
Denat: 82 °C
:05
Present Temp: 45 °C
PGM 01 HER2
Denat in Process
Denat: 82 °C 02:28
Present Temp: 82 °C
Please Wait
Cooling to Hyb 45 °C
Present Temp: 58 °C
PGM 01 HER2
Hyb in Process
Hyb: 45 °C
12:48
Present Temp: 45°C
PGM 01 HER2
PROCESS COMPLETE
Total Hyb Time 20:00
End PGM/Main Menu