This procedure should be carried out at least three times or until an acceptable

agreement between the measured volumes and the digital displays are

achieved.

Note: You should only attempt to calibrate in one direction.

Where more than 100cc of volume change occurs, you will need to reverse the

flow. You must therefore check the reverse displacement against the calibrated

readings. If necessary you may need to use a correction factor to overcome any

small hysteresis.