beautypg.com

Humboldt HM-2315 Volume Change Apparatus User Manual

Page 3

background image

Volume Change Measurement & Bypass
In order to measure the volume change, the left hand valve of the Volume

Change Apparatus must be set to VOLUME CHANGE position and the right

hand valve must be set to FLOW UP position or FLOW DOWN position.

During a test, water fills one of the chambers, while the other chamber is

opened and water drains out. As one chamber is filling and the other draining,

the piston will move inside the cylinder and the lever arm will move up or

down depending on which chamber is filling. The lever arm acts on the LSCT

or Digital Gauge which indicates the distance the piston has moved. This

information is then converted by the computer to volume change. During the

test, if the capacity of the sample exceeds the volume in one of the chambers,

it may be necessary to increase the volume change capacity. This is done by

reversing the flow direction. The flow direction is changed by switching the

FLOW UP/FLOW DOWN valve to the opposite direction. Note: When using

the reverse flow for greater capacity during a test, certain precautions should be

carried out when calibrating the apparatus (see calibration instructions).

By setting both valves to the BYPASS position, de-aired water can flow

through the change over valves system direct to the Triaxial cell. Volume is not

measured when the valves are in the Bypass position.

Calibration:
Switch on the power supply to the logger or monitoring readout unit with the

LSCT or Digital Gauge connected to it for at least one hour before attempting

calibration. Note: Some older monitors or loggers require a longer warm up

period. Please refer to your logger or read out unit guide.

With the probe of the transducer or Digital Gauge set to its maximum

downward position, set the monitor or logger to zero (if using a Digital Gauge,

set it to zero also). Note: The probe must still be in contact with the lever arm.

Adjust the right hand valve to the FLOW UP position until the maximum upward

movement has been achieved. Accurately measure the amount of water flowing

out with a measuring cylinder.

Alternatively you could work out the volume change by weighing the volume of

water displaced from the draining chamber. Weigh a beaker or other suitable

container with which to collect the water. Collect the water displaced when one

chamber is filled and the other drained. Weigh the beaker again with the water

displaced after the maximum upward movement has been achieved. Subtract

the empty container weight from the weight of the container and displaced

water. If the weight is measured in grams, this number will be the same as the

volume in millilitres or cubic centimetres.

Once the volume has been determined, the Digital Gauge should be set to

show the volume displaced and the logger or monitor should be adjusted to the

same volume.