Place the specimen in position by setting one end of the specimen mold into

one of the three mounting pins on the end plate. Set the other end of the mold

into the corresponding pin on the movable carriage.

Move the length indicator to the zero position on the scale located on the upper

side of the tank by loosening the adjusting screw. The length indicator will then

read the distance the specimen has been pulled apart in centimeters.

At the completion of the test return the carriage to the starting position.

Note: If the test material comes in contact with the surface of the water or the

bottom of the bath, the test is not considered satisfactory. The specific gravity

of the bath may be adjusted by the addition of either methyl alcohol or sodium

chloride to prevent the bituminous sample from coming to the surface of the

water or touching the bottom of the bath during the test.

Operating Instructions
The Temperature Controlled Ductility Machine has a separate thermostatically

controlled constant temperature circulator with recommended fluids circulating

through finned tubes in the trough. Recommended circulator fluids are 50/50

filtered or distilled water and glycol to maintain a constant temperature of within

± 0.9°F (0.5°C).

Test Report
The test is considered normal when the rupture occurs at the point where the

thread has practically no cross-sectional area.

The average of three normal tests is reported as the ductility of the sample.

Maintenance
All bearing surfaces should be kept clean. the oil cups provide for adding SAE

Grade 30 lubricating oil. Threaded lead screw should be kept clean and properly

greased to ensure proper carriage level travel. Use NGLI #2 Extreme Pressure

Lithium Grease or equivalent on the threads. When not in use, remove line cord

from the outlet to prevent misuse or improper handling.