Procedures

A) Equipment Check

1) Before beginning a test, check to ensure the Drive Rod is straight by

rolling the rod on a flat surface. Drive Rod may have been bent if

overloaded vertically.

2) The Drive Cone Assembly must be checked to ensure proper movement

and that the point is discernable. The point may eventually become

rounded depending on the amount of use and types of material tested.

The original diameter of the cone is 0.544". Replacement should be

considered if the diameter is reduced to 0.528" or if substantial damage

occurs to the cone point.

B) Assembling
1)

Starter Rod:

Insert starter rod in bottom of head assembly and tighten

with 7/16" wrench.

1)

Extension Rod (if used):

Attach extension rod either above or below

starter rod. (Inside rod on starter rod is shorter than that of extension

rod.) Tighten with 7/16" wrench.

2)

Drive Cone Assy.:

Attach to lower section and securely tighten with

7/16" wrench.

a) The standard 1.5 cm

2

Drive Cone Assy. is used in typical low

strength fine grained soft soils.

b) The optional 3 cm

2

Drive Cone Assy. is used in very soft, typically

uncompacted materials where readings in psf may fall below 2000

psf. Use of this cone requires that the displayed reading be divided

by 2 for the correction.

C) Testing Sequence
1)

Tare Display:

This may be done by setting the DSCP on horizontal

surface and turning on while allowing weight of head to rest on cone.

Unit will tare automatically upon startup. Tare function may also be

achieved at any time while unit is on by allowing the unit to stand on

horizontal surface while allowing weight of head to rest on cone and

pressing the “Tare” button.

2)

Select Units:

Units desired for test may be selected by holding down

“MAX” button and at the same time pushing the “ZERO” button

which will scroll thru the units available.

4