COVIDIEN Genius™ 2 Tympanic Thermometer User Manual
Page 13

English
Tympanic Thermometer and Base
10
Calibrated Accuracy Limits (after Recalibration*):
Ambient Temperature
Target Temperature
Accuracy
16°C to 33°C
36°C to 39°C
± 0.2°C
(60.8°F to 91.4°F)
(96.8°F to 102.2°F)
(± 0.4°F)
16°C to 33°C
less than 36°C
± 0.3°C
(60.8°F to 91.4°F)
(less than 96.8°F)
(± 0.5°F)
16°C to 33°C
greater than 39°C
± 0.3°C
(60.8°F to 91.4°F)
(greater than 102.2°F)
(± 0.5°F)
*Post recalibration accuracy using the Genius 2 Checker/Calibrator may not necessarily be equivalent to factory calibration.
Displayed Temperature Measurement Range:
Temperature Range depends on Site Mode as follows:
Mode
Range °C
Range °F
Ear
33.0 to 42.0
91.4 to 107.6
Oral
33.6 to 42.0
92.5 to 107.6
Core
34.0 to 42.0
93.2 to 107.6
Rectal
34.2 to 42.0
93.6 to 107.6
Ambient Temperature Range:
16°C to 33°C (60.8°F to 91.4°F) where as CEN defines the standard as 16°C to 35°C (60.8°F to 95°F), 10 to 95 %RH, non-condensing.
Storage Temperature Range:
-25°C to 55°C (-13°F to 131°F), up to 95% RH non-condensing. If the unit is stored at extremes, it is recommended that the unit be
checked on the field calibration checker or at the factory before returning to service.
Clinical Repeatability:
Meets Section A.5 of EN 12470-5: 2003 (E) per Covidien llc technical report. Data is available from Covidien llc on request.
Response Time:
Less than 2 seconds
Pulse Timer:
60 seconds
Temperature Resolution:
0.1°C or 0.1°F
Power:
3 AAA alkaline batteries
Battery Life:
Minimum of 15,000 temperature readings
Size:
Thermometer — 17.8 cm (7”)
Base — 20.3 cm (8”)
Weight:
Thermometer (with batteries) — 160 grams
Base — 100 grams