Section I - Functional Description
This manual describes the operation of the Genius™ 2 Checker/Calibrator. This device has been developed to check the accuracy
of the Genius™ 2 Tympanic Thermometer and automatically recalibrate the thermometer, if necessary. In addition, the Genius 2
Checker/Calibrator has the capability to generate a test report for each thermometer tested and save it to a USB flash drive.
The Genius 2 Checker/Calibrator will work with all Genius™ 2 Thermometer software revisions.
The Genius 2 Checker/Calibrator contains two independently controlled infrared calibration targets that are similar to factory
calibration targets. These “blackbody” targets are designed to have efficient radiative heat transfer and produce infrared radiation
that approximates the human ear at temperatures of 32.2° C (90.0° F ) and 40.6° C (105.0° F ).
F/C˚
F/C˚
1
2
3
4
8
9
10
5
6
7
11
F 3.15A 250V
Checker / Calibrator
Figure 1 - Front Panel Layout
Legend
1 Power Connector
7 High Temperature Target
2 Power Switch
8 High Temperature Target LED
3 Fuse Holder
9 LCD Display
4 Low Temperature Target LED
10 Encoder Knob
5 Low Temperature Target
11 USB port