COVIDIEN Genius™ 2 Tympanic Thermometer Checker-Calibrator User Manual
Page 14

Genius
TM
2
Checker/Calibrator
Table of Contents
12
Section XIII - Electromagnetic Conformity Declaration
The Genius 2 Checker/Calibrator has been built and tested according to IEC/EN61010-1 and EN61326-1 standards
The Checker/Calibrator is intended for use in the electromagnetic environment specified below. The user of the equipment should
assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment -Guidance
Conducted and Radiated
Emissions (EN61326-1)
Group 1, Class B
The Checker / Calibrator emissions are very low and are not likely
to cause interference in nearby electronic equipment.
Harmonic Current
(EN61000-3-2)
Class A
Voltage Fluctuation and Flicker
(EN61000-3-3)
Complies
Radiated Disturbance Immunity
(EN61000-4-3)
Complies
Conducted Disturbance
Immunity (EN61000-4-6)
Complies
Power Frequency Magnetic Field
Immunity (EN61000-4-8)
Complies
Voltage Dips and Interrupts
Immunity (EN61000-4-11)
Complies
Electrical Fast Transient/Bursts
Immunity (EN61000-4-4)
Complies
Surge Immunity (EN61000-4-5)
Complies
Electrostatic Discharge
(EN61000-4-2)
Meets Criteria C,
Non-Continuous