COVIDIEN Argyle™ Peripherally Inserted Central Catheter User Manual

Page 2

background image

2

n

en

DESCRIPTION:
A radiopaque, polyurethane Peripherally Inserted Central Catheter (PICC)
INDICATIONS:
The catheter is designed for cases in which venous catheterization or long term I.V. administration is

necessary. Placement is routinely achieved from a peripheral venous site, but the catheter may be

placed via subclavian cutdown as well. The catheter may be used to administer fluids for hydration and

parenteral nutrition, as well as other commonly used intravenous medications.
Note:
Before performing catheter placement all directions, precautions, contraindications, and

warnings must be read and clearly understood. A standard protocol should be established

prior to use of the Argyle™ PICC, and the catheter is only to be placed by personnel with

appropriate training. The CDC (Centers for Disease Control) guidelines can be consulted

to assist with the development of protocols for the care and use of peripherally inserted

venous catheters.

CATHETER SPECIFICATIONS (NOMINAL):

Reorder #

Description

French

Size

(Fr)

EffectiveLength

(cm)

OuterDiameter

(mm)

Inner

Diameter

(mm)

43303

Peripheral Inserted

Central Catheter (PICC)

Single Lumen

1.9

30

0.6

0.4

43304

Peripheral Inserted

Central Catheter (PICC)

Dual Lumen

1.9

30

0.6

N/A1

1

Dual Lumen Primary and Secondary lumen areas are the same and “D” shaped.

Dual Lumen PICC

Single Lumen PICC

Primary Lumen

Secondary Lumen

Length

Trimmed

(cm)

Remaining

Effective

Length (cm)

Priming

Volume

(mL)

Flow Rate 100

cm Gravity

(mL/hr)

Priming

Volume

(mL)

Flow Rate 100

cm Gravity

(mL/hr)

Priming

Volume

(mL)

Flow Rate

100 cm

Gravity

(mL/hr)

0

30

0.18

55

0.15

6

0.13

6

4

26

0.15

63

0.14

7

0.12

7

8

22

0.13

71

0.13

8

0.11

8

12

18

0.12

81

0.12

10

0.10

10

16

14

0.12

91

0.12

12

0.10

12

20

10

0.12

114

0.12

16

0.10

16

• The dual lumen PICC is capable of infusing a 20% IV fat emulsion at 50 mL/Hr with less than 12 psi

infusion pump pressure.

• Graduation marks are printed every cm starting at 3cm at the distal end and terminating at 30cm at

the proximal end near the stabilizing wing. The 5cm marks are bold with the 20cm and 30cm marks

identified with two and three bold marks respectively.

• At the time of manufacture the distal tip is formed with a straight or blunt cut configuration.

PRECAUTIONS, CONTRAINDICATIONS AND WARNINGS:

• Do not use if unit package is opened or damaged.
• Use of aseptic technique and Universal Precautions should be followed during the placement and

maintenance of the catheter.

• To ensure the catheter does not migrate, and to be certain it is not inadvertently broken, it must be

secured according to the instructions.

• Do not place the catheter tip or allow the catheter tip to migrate into the low right atrium. This