beautypg.com

Cooper Instruments & Systems DTW 400 Digital Torque Tester / Screwdriver User Manual

Page 5

background image

CF132 DTW 400/405

5

11/12

Output is available in PD, PP and C modes. In PD and PP modes, peak data is output when ZERO is pressed
or activated by the AUTO ZERO function. In C mode, the gauge outputs data continuously 160 data / second.

Comport signal

8 data, 2 stop, no parity. Baud rate: 19,200 bps


Output connector

USB A/B cable


Peak Data Output Format


[CAN] _ _ _ [SO] [value] _ [SI] [unit] [CR]

( _ _ _ is memory location)

[CAN]:

ASCII control code 24

_:

Space (code 32)

[SO]:

ASCII control code 14

[value]:

Output data with sign and decimal point. Plus sign represents CW torque and minus
sign for CCW. [value] always occupies six locations and empty locations will be filled
with spaces.

[SI]:

ASCII control code 15

[unit]:

N*m _ _ _=N·m

kgf*cm=kg·cm
lb*in=lb·in
[CR]:

ASCII control code 13 (Carriage Return)

Continuous Output Data Format

[CAN] [value] [CR]

Changing the Continuous Output Data Rate

1. Turn the tester off.
2. Click ON/OFF while holding PRG, “00” is displayed.
3. Press UP to select either “00” or “01”.

00 = 12 data/second
01 = 180 data/second

4. Press PRG to save and exit.