To Configure Scaling Factor:

1.

Press

/MIN to display and to select

the digit (or decimal point) you want
to change.

2.

Press

/MAX to increase the value

of the selected digit.

3.

Repeat steps 1 and 2 until each digit is the desired value
(your calculated scaling factor).

4.

Press MENU.
The meter displays:

To Configure Meter Inputs:

1.

Press MENU until
the meter displays:

2.

Press

/MIN to display:

3.

Press

/MAX, if necessary to change the configuration

value.

4.

Repeat steps 2 and 3 for

INP.2

through

INP.7

5.

Press MENU.
The meter displays:

SETPTS

/MIN

MENU

RESET

/MAX

in.sc.of

SETPTS

/MIN

MENU

RESET

/MAX

stored

SETPTS

/MIN

MENU

RESET

/MAX

inp.1=0

SETPTS

/MIN

MENU

RESET

/MAX

in cnf

SETPTS

/MIN

MENU

RESET

/MAX

rdg of

SETPTS

/MIN

MENU

RESET

/MAX

stored

SETPTS

/MIN

MENU

RESET

/MAX

0.00000

To Configure Type of Input:

1.

Press MENU until the
meter displays:

2.

Press

/MIN to display a flashing input type.

3.

Press MENU.

The meter displays:

4.

Press MENU and

bridge

stops flashing.

5.

Press MENU.

The meter displays:

To Configure Meter Display Readings:

1.

Press

/MIN to display:

2.

Press

/MAX, if necessary to change the configuration

value to

0

or

1

.

3.

Repeat steps 1 and 2 for

RDG.2

through

RDG.7

4.

Press MENU.

The meter displays:

SETPTS

/MIN

MENU

RESET

/MAX

rdg SC

SETPTS

/MIN

MENU

RESET

/MAX

stored

SETPTS

/MIN

MENU

RESET

/MAX

rdg.1=0

SETPTS

/MIN

MENU

RESET

/MAX

rdg.cnf

SETPTS

/MIN

MENU

RESET

/MAX

stored

SETPTS

/MIN

MENU

RESET

/MAX

bridge

SETPTS

/MIN

MENU

RESET

/MAX

stored

SETPTS

/MIN

MENU

RESET

/MAX

input1

To Set Decimal Point Position:

1.

Press MENU until
the meter displays:

2.

Press

/MIN to display:

Press

/MAX to move

the decimal point to the
desired location.

3.

Press MENU.
The meter displays:

4.

Press RESET twice. The meter flashes
and displays an output reading.

The meter is now in RUN mode.

Configure Reading Offset

Now that you are in the run mode with a transducer connected
to the meter, do the following:

1.

Simulate a load on the transducer (leave the pressure
port open).

2.

Note the display reading. LetÕs assume
the display shows 43.5.

3.

To make the display show zeroes, press
MENU until the meter displays:

4.

Press

/MIN to display the previous reading offset value.

5.

Using

/MIN to scroll through the digits

and

/MAX to change the value, enter

the value -0043.5.

6.

Press MENU.
The meter displays:

SETPTS

/MIN

MENU

RESET

/MAX

reset2

SETPTS

/MIN

MENU

RESET

/MAX

cntby

SETPTS

/MIN

MENU

RESET

/MAX

stored

SETPTS

/MIN

MENU

RESET

/MAX

FFFF.FF

SETPTS

/MIN

MENU

RESET

/MAX

DeC pt

7.

Press RESET twice.
The meter flashes and
then displays a value.

Your meter is now in RUN mode and operational.

If You Have Bipolar Input ±50mV

The typical setting for your meter is unipolar. If, however, you
have bipolar input ±50mV, you must install jumper S1B.
Remove the outer panel mounting sleeve to expose the
jumper.

S1 Jumper Location on Signal Input Board

In addition, you must set configuration menu value INP.3=1
(under IN CNF menu). Refer to the Configuration sections of
this Quick Start manual.

A

B

A

D

G

H

V

L

P

S

T

U

B

C

E

F

J

K

N

M

R

Q

S2

J3

J9

S1

SETPTS

/MIN

MENU

RESET

/MAX

reset2

SETPTS

/MIN

MENU

RESET

/MAX

stored

SETPTS

/MIN

MENU

RESET

/MAX

in cnf

SETPTS

/MIN

MENU

RESET

/MAX

rdg of

SETPTS

/MIN

MENU

RESET

/MAX

-0043.5