In.sc.of, Stored, Inp.1=0 – Cooper Instruments & Systems DFI INFINITY Digital Force Indicator/Controller User Manual
Page 2: Rdg of, Rdg sc, Rdg.1=0, Rdg.cnf, Bridge, Input1 to set decimal point position, Configure reading offset

5
6
7
8
To Configure Scaling Factor:
1.
Press
ᮣ
/MIN to display and to select
the digit (or decimal point) you want
to change.
2.
Press
ᮡ
/MAX to increase the value
of the selected digit.
3.
Repeat steps 1 and 2 until each digit is the desired value
(your calculated scaling factor).
4.
Press MENU.
The meter displays:
To Configure Meter Inputs:
1.
Press MENU until
the meter displays:
2.
Press
ᮣ
/MIN to display:
3.
Press
ᮡ
/MAX, if necessary to change the configuration
value.
4.
Repeat steps 2 and 3 for
INP.2
through
INP.7
5.
Press MENU.
The meter displays:
ᮡ
ᮣ
SETPTS
/MIN
MENU
RESET
/MAX
in.sc.of
ᮡ
ᮣ
SETPTS
/MIN
MENU
RESET
/MAX
stored
ᮡ
ᮣ
SETPTS
/MIN
MENU
RESET
/MAX
inp.1=0
ᮡ
ᮣ
SETPTS
/MIN
MENU
RESET
/MAX
in cnf
ᮡ
ᮣ
SETPTS
/MIN
MENU
RESET
/MAX
rdg of
ᮡ
ᮣ
SETPTS
/MIN
MENU
RESET
/MAX
stored
ᮡ
ᮣ
SETPTS
/MIN
MENU
RESET
/MAX
0.00000
To Configure Type of Input:
1.
Press MENU until the
meter displays:
2.
Press
ᮣ
/MIN to display a flashing input type.
3.
Press MENU.
The meter displays:
4.
Press MENU and
bridge
stops flashing.
5.
Press MENU.
The meter displays:
To Configure Meter Display Readings:
1.
Press
ᮣ
/MIN to display:
2.
Press
ᮡ
/MAX, if necessary to change the configuration
value to
0
or
1
.
3.
Repeat steps 1 and 2 for
RDG.2
through
RDG.7
4.
Press MENU.
The meter displays:
ᮡ
ᮣ
SETPTS
/MIN
MENU
RESET
/MAX
rdg SC
ᮡ
ᮣ
SETPTS
/MIN
MENU
RESET
/MAX
stored
ᮡ
ᮣ
SETPTS
/MIN
MENU
RESET
/MAX
rdg.1=0
ᮡ
ᮣ
SETPTS
/MIN
MENU
RESET
/MAX
rdg.cnf
ᮡ
ᮣ
SETPTS
/MIN
MENU
RESET
/MAX
stored
ᮡ
ᮣ
SETPTS
/MIN
MENU
RESET
/MAX
bridge
ᮡ
ᮣ
SETPTS
/MIN
MENU
RESET
/MAX
stored
ᮡ
ᮣ
SETPTS
/MIN
MENU
RESET
/MAX
input1
To Set Decimal Point Position:
1.
Press MENU until
the meter displays:
2.
Press
ᮣ
/MIN to display:
Press
ᮡ
/MAX to move
the decimal point to the
desired location.
3.
Press MENU.
The meter displays:
4.
Press RESET twice. The meter flashes
and displays an output reading.
The meter is now in RUN mode.
Configure Reading Offset
Now that you are in the run mode with a transducer connected
to the meter, do the following:
1.
Simulate a load on the transducer (leave the pressure
port open).
2.
Note the display reading. LetÕs assume
the display shows 43.5.
3.
To make the display show zeroes, press
MENU until the meter displays:
4.
Press
ᮣ
/MIN to display the previous reading offset value.
5.
Using
ᮣ
/MIN to scroll through the digits
and
ᮡ
/MAX to change the value, enter
the value -0043.5.
6.
Press MENU.
The meter displays:
ᮡ
ᮣ
SETPTS
/MIN
MENU
RESET
/MAX
reset2
ᮡ
ᮣ
SETPTS
/MIN
MENU
RESET
/MAX
cntby
ᮡ
ᮣ
SETPTS
/MIN
MENU
RESET
/MAX
stored
ᮡ
ᮣ
SETPTS
/MIN
MENU
RESET
/MAX
FFFF.FF
ᮡ
ᮣ
SETPTS
/MIN
MENU
RESET
/MAX
DeC pt
7.
Press RESET twice.
The meter flashes and
then displays a value.
Your meter is now in RUN mode and operational.
If You Have Bipolar Input ±50mV
The typical setting for your meter is unipolar. If, however, you
have bipolar input ±50mV, you must install jumper S1B.
Remove the outer panel mounting sleeve to expose the
jumper.
S1 Jumper Location on Signal Input Board
In addition, you must set configuration menu value INP.3=1
(under IN CNF menu). Refer to the Configuration sections of
this Quick Start manual.
A
B
A
D
G
H
V
L
P
S
T
U
B
C
E
F
J
K
N
M
R
Q
S2
J3
J9
S1
ᮡ
ᮣ
SETPTS
/MIN
MENU
RESET
/MAX
reset2
ᮡ
ᮣ
SETPTS
/MIN
MENU
RESET
/MAX
stored
ᮡ
ᮣ
SETPTS
/MIN
MENU
RESET
/MAX
in cnf
ᮡ
ᮣ
SETPTS
/MIN
MENU
RESET
/MAX
rdg of
ᮡ
ᮣ
SETPTS
/MIN
MENU
RESET
/MAX
-0043.5