Step 1:
Remove all secondary check parts from the
existing outlets. Replace the NCG rubber seat with
Amico #15 O-Ring.
Retro-Fit Instructions
Step 2:
Thread the Amico Retro-Fit rough-in assembly to
the existing NCG rough-in.
Important Note
Due to unknown factors of the existing equipment, projects requiring Retro-Fit Conversions must be fully investigated by
the installer.
It is the responsibility of the installer to determine if the Retro-Fit will function correctly and if additional components are
required. Amico is not responsible for any additional components required.
O-VC-OH1-TOOL
Amico Tool
Hole for Screw Driver
(to loosen & tighten)
Existing NCG
Rough-In
O-Ring
(H-ORING-015VTN)
NCG Original Rubber
Seat (to be removed &
replaced by #15 O-Ring)
Existing NCG
Rough-In
Back Rough-In
Assembly Kit
(O-VC-NCG-U-GAS)
- Artwall (6 pages)
- Ceiling Column Manual Retractable (16 pages)
- Ceiling Column Motorized Retractable (20 pages)
- Ceiling Column Rigid (Stationary) (12 pages)
- Ceiling Column Rigid LITE (Stationary) (12 pages)
- Recessed Console Unit (8 pages)
- Freedom Series Recessed (12 pages)
- Freedom Series Surface Mounted (8 pages)
- LDR Headwall (8 pages)
- Majestic Series Double Tier Recessed Headwall (6 pages)
- Majestic Series Double Tier Surface Mounted Headwall (8 pages)
- Majestic Series Single Tier Surface Mounted Headwall (6 pages)
- Majestic Series Single Tier Surface Mounted Headwall (8 pages)
- Majestic Series Single Tier Recessed Headwall (8 pages)
- Noble Custom Cabinet (7 pages)
- Power Column (7 pages)
- Regal Series Recessed Vertical Flatwall (8 pages)
- Regal Series Surface Mounted Vertical Flatwall (7 pages)
- Security Recessed Console Box Assembly (12 pages)
- Surface Mount Console Unit (6 pages)
- Hydraulic Patient Transfer Stretcher (S-H-300) (20 pages)
- Hydraulic Patient Transfer Stretcher (S-H-300) (11 pages)
- AHM Monitor Arms (Adjustable Height) (34 pages)
- Barcode Dock Mount (15 pages)
- Condor Adjustable Height Channel (20 pages)
- Condor Lite Fixed Height Channel (31 pages)
- Condor Lite Fixed Height Channel (12 pages)
- Falcon Combo Arm (44 pages)
- Light-Duty Monitor Arm (20 pages)
- Monitor Roll Stands (12 pages)
- CSA Duplex RED Horizontal Tank Mount (36 pages)
- NFPA Duplex CCD Horizontal Tank Mount (51 pages)
- NFPA Duplex RED Horizontal Tank Mount (40 pages)
- 12" Floor Pedestal (8 pages)
- 2 in 1 Alarm System (56 pages)
- Alarm Valve Combo Unit (32 pages)
- Alert-3 LCD Alarm (28 pages)
- Automatic Cryogenic Liquid Solenoid Manifold (32 pages)
- Ball Valve (8 pages)
- British Standard Outlet (18 pages)
- British Standard Outlet Slim (18 pages)
- Chemetron Compatible Ceiling Column Outlet (24 pages)
- Chemetron Retro-fit Gas Control Panel (4 pages)
- Dome Loaded Manifold CSA (44 pages)
- Dome Loaded Manifold Nema-4 NFPA (44 pages)