Amico Medical Gas Check Valve with Extensions User Manual
Page 4

12. Test the system (per appropriate standards) to ensure that the intended gas service is
connected to the appropriate line.
13. Test the system for leaks (per appropriate standards).
14. If no leaks are found, gauges (purchased separately) can be installed, if required. Pipe
sealants should comply with requirements of NFPA 99 or CSA Z7396.1. Use care to exclude
pipe sealants from the check valve cavity and from interior tube surfaces exposed to medical
gas flow service. Properly applied Teflon tape is an alternative to pipe sealants.
4
Amico Pipeline
See also other documents in the category Amico Equipment:
- Artwall (6 pages)
- Ceiling Column Manual Retractable (16 pages)
- Ceiling Column Motorized Retractable (20 pages)
- Ceiling Column Rigid (Stationary) (12 pages)
- Ceiling Column Rigid LITE (Stationary) (12 pages)
- Recessed Console Unit (8 pages)
- Freedom Series Recessed (12 pages)
- Freedom Series Surface Mounted (8 pages)
- LDR Headwall (8 pages)
- Majestic Series Double Tier Recessed Headwall (6 pages)
- Majestic Series Double Tier Surface Mounted Headwall (8 pages)
- Majestic Series Single Tier Surface Mounted Headwall (6 pages)
- Majestic Series Single Tier Surface Mounted Headwall (8 pages)
- Majestic Series Single Tier Recessed Headwall (8 pages)
- Noble Custom Cabinet (7 pages)
- Power Column (7 pages)
- Regal Series Recessed Vertical Flatwall (8 pages)
- Regal Series Surface Mounted Vertical Flatwall (7 pages)
- Security Recessed Console Box Assembly (12 pages)
- Surface Mount Console Unit (6 pages)
- Hydraulic Patient Transfer Stretcher (S-H-300) (20 pages)
- Hydraulic Patient Transfer Stretcher (S-H-300) (11 pages)
- AHM Monitor Arms (Adjustable Height) (34 pages)
- Barcode Dock Mount (15 pages)
- Condor Adjustable Height Channel (20 pages)
- Condor Lite Fixed Height Channel (31 pages)
- Condor Lite Fixed Height Channel (12 pages)
- Falcon Combo Arm (44 pages)
- Light-Duty Monitor Arm (20 pages)
- Monitor Roll Stands (12 pages)
- CSA Duplex RED Horizontal Tank Mount (36 pages)
- NFPA Duplex CCD Horizontal Tank Mount (51 pages)
- NFPA Duplex RED Horizontal Tank Mount (40 pages)
- 12" Floor Pedestal (8 pages)
- 2 in 1 Alarm System (56 pages)
- Alarm Valve Combo Unit (32 pages)
- Alert-3 LCD Alarm (28 pages)
- Automatic Cryogenic Liquid Solenoid Manifold (32 pages)
- Ball Valve (8 pages)
- British Standard Outlet (18 pages)
- British Standard Outlet Slim (18 pages)
- Chemetron Compatible Ceiling Column Outlet (24 pages)
- Chemetron Retro-fit Gas Control Panel (4 pages)
- Dome Loaded Manifold CSA (44 pages)
- Dome Loaded Manifold Nema-4 NFPA (44 pages)