Maintenance
1. If a problem should arise, do not disassemble the valve while the line is under pressure.
2. Repair kits are available, should the check valve need replacing.
Operation
1. No manual operation is required, of the check valve, once it is in the pipe line.
NOTE: Not recommended for use with reciprocating pumps and similar applications which
may induce repetitious vibrations. Low flow rates, which do not fully open the valve, may
result in undesirable noise and premature valve failure. Upstream flow disturbances, which
create turbulence, may also result in rapid wear. Therefore it is recommended that a minimum
of 10 diameters of straight pipe be provided between the check valve and any upstream flow
disturbances such as pumps, control valves, elbows, etc.
Caution
It is the installers and/or system designers responsibility to ensure that these valves
are installed in accordance with applicable and current NFPA/CSA Standards.
A m i c o M i c r o p r o c e s s o r B a s e d A l a r m
P a g e : 4
USER RESPONSIBILITY
The information contained in this Installation and Operation Maintenance
Manual, pertains only to the ALERT-2 microprocessor based digital alarm.
This product will perform to conformity with the descriptions contained in this
manual, when assembled, operated, maintained and serviced in accordance
with the installation instructions provided.
The alarm
must be checked periodically. Parts that are broken, missing,
worn, distorted or contaminated,
must be replaced immediately. Should such
repair or replacement become necessary, please contact Amico Corporation
or their distributors.
All alarms should not be repaired, or altered without prior written or verbal
approval of Amico Corporation or it’s distributors. Failure to comply will void
all warranty on the alarm.
Statements in this manual preceded by the words
WARNING
,
CAUTION
,
DANGER and
NOTE
are of special significance. Please read these sections
carefully.
WARNING:
denotes steps which can prevent injury.
CAUTION:
denotes steps which can prevent damage to equipment.
DANGER:
denotes steps which can prevent electrical shock to equipment
or to prevent serious injury and/or death.
2