Control board replacement – Amico Dome Loaded Manifold CSA User Manual
Page 28
28
Amico Pipeline
Control Board Replacement
If necessary, the control board replacement can be performed while the manifold is in service. However, this should only
be done by qualified technicians experienced in servicing medical equipment.
1. Remove two top screws at the base of the display board bracket and tilt the display board assembly backwards.
2. Disconnect power supply.
3. Remove the screws located on the circuit board, which detaches it from the front plate. Note orientation of the
old circuit board and remove the wiring one at a time and connect each wire to the appropriate terminal on the
new board.
4. Install the circuit board and fasten the screws to secure it. Tilt the circuit board to its original position and then
fasten the two screws at the base.
5. Reconnect power supply to the manifold cabinet.
6. Check the correct operation of the circuit board by simulating a bank switch over a couple of times.
- Artwall (6 pages)
- Ceiling Column Manual Retractable (16 pages)
- Ceiling Column Motorized Retractable (20 pages)
- Ceiling Column Rigid (Stationary) (12 pages)
- Ceiling Column Rigid LITE (Stationary) (12 pages)
- Recessed Console Unit (8 pages)
- Freedom Series Recessed (12 pages)
- Freedom Series Surface Mounted (8 pages)
- LDR Headwall (8 pages)
- Majestic Series Double Tier Recessed Headwall (6 pages)
- Majestic Series Double Tier Surface Mounted Headwall (8 pages)
- Majestic Series Single Tier Surface Mounted Headwall (6 pages)
- Majestic Series Single Tier Surface Mounted Headwall (8 pages)
- Majestic Series Single Tier Recessed Headwall (8 pages)
- Noble Custom Cabinet (7 pages)
- Power Column (7 pages)
- Regal Series Recessed Vertical Flatwall (8 pages)
- Regal Series Surface Mounted Vertical Flatwall (7 pages)
- Security Recessed Console Box Assembly (12 pages)
- Surface Mount Console Unit (6 pages)
- Hydraulic Patient Transfer Stretcher (S-H-300) (11 pages)
- Hydraulic Patient Transfer Stretcher (S-H-300) (20 pages)
- AHM Monitor Arms (Adjustable Height) (34 pages)
- Barcode Dock Mount (15 pages)
- Condor Adjustable Height Channel (20 pages)
- Condor Lite Fixed Height Channel (31 pages)
- Condor Lite Fixed Height Channel (12 pages)
- Falcon Combo Arm (44 pages)
- Light-Duty Monitor Arm (20 pages)
- Monitor Roll Stands (12 pages)
- CSA Duplex RED Horizontal Tank Mount (36 pages)
- NFPA Duplex CCD Horizontal Tank Mount (51 pages)
- NFPA Duplex RED Horizontal Tank Mount (40 pages)
- 12" Floor Pedestal (8 pages)
- 2 in 1 Alarm System (56 pages)
- Alarm Valve Combo Unit (32 pages)
- Alert-3 LCD Alarm (28 pages)
- Automatic Cryogenic Liquid Solenoid Manifold (32 pages)
- Ball Valve (8 pages)
- British Standard Outlet (18 pages)
- British Standard Outlet Slim (18 pages)
- Chemetron Compatible Ceiling Column Outlet (24 pages)
- Chemetron Retro-fit Gas Control Panel (4 pages)
- Dome Loaded Manifold Nema-4 NFPA (44 pages)