Initial power-up – Amico Dome Loaded Manifold CSA User Manual
Page 19

www.amico.com
19
Initial Power-Up
Release the two latches on either side of the manifold control panel and remove the cover.
Verify the following:
• Both master valves located on both header bars are turned fully counter clockwise (open).
• All four line regulator isolation valves are open (handles horizontal).
• Power supply has been connected.
• Both Red “empty” indicators on the front of the manifold are illuminated.
• If connected to a master alarm panel, “Secondary Supply” alarm is activated.
Close 3/4” source shut-off valve.
“S-L-O-W-L-Y” open one cylinder on the right side of the manifold.
Verify the following:
• Right bank Red “Empty” LED goes out.
• Right bank Green “in use” LED illuminates.
• Right bank cylinder contents gauge reads cylinder pressure.
“S-L-O-W-L-Y” open one cylinder on the left side of the manifold.
Verify the following:
• Left bank Red “Empty” LED goes out.
• Left bank Yellow “Ready” LED illuminates.
• Left bank cylinder contents gauge reads cylinder pressure.
• If connected to a master alarm panel, “secondary supply” alarm is not activated.
Close right bank cylinder. Slightly depress bleed valve located on the side of the regulator. Verify the
following:
• Right bank cylinder contents gauge drops “S-L-O-W-L-Y”.
• As the right cylinder contents gauge is nearly depleted, the manifold changes over to the left bank.
• After change-over, the right bank Green “In Use” LED goes out and the Red “Empty” LED illuminates.
• After change-over, the left bank Yellow “Ready” LED goes out and the Green “In Use” LED illuminates.
Verify the “line pressure” gauge reading is acceptable.
“S-L-O-W-L-Y” open one cylinder on the right side of the manifold.
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.