beautypg.com

Inspection and testing installation and dimensions – Amico British Standard Outlet User Manual

Page 4

background image

4

Amico Pipeline

Inspection and Testing

Installation and Dimensions

Pressure drops across the Amico British Outlet shall comply with clause 6.3 of BS EN 737-1:1998.

Note:

The Amico British Outlets meet and exceed the requirements at the time of manufacture. However piping source

capacity, sizing and restrictions may prevent outlets form attaining these values.

The Amico British Outlet should be mounted at a height of between 900 mm and 1400 mm above finished floor level

and not less than 200 mm from any obstruction.

Where more than one Outlet unit is to be mounted in one location, these should be mounted at the following spacing;

• Two Outlet units - 150 mm centres
• Three or more Outlets units - 135 mm centres

Outlet units mounted in a horizontal array shall be installed in the following sequence when viewed from the front, left

to right;

O₂; N₂O; 50%O₂/50%N₂O; Medical Air, Surgical Air, Vacuum; AGSS

Please refer to HTM 2022, section 3.

AGSS Requirements for flow and pressure drop across terminal units with probe inserted clause 5.4.8 of BS EN 737-4:1998

Nominal Pressure

(kPa)

Test Pressure

(kPa)

Test Flow

(1/min)

EN 737-1 Limit

(kPa)

400

320

60

15

400

320

200

70

700

640

300

70

Vacuum

*40

40

15

* Absolute pressure

Terminal Type

Test Pressure

Test Flow

(1/min)

Maximum Pressure Drop

Across a Terminal Unit

(kPa)

1

Atmospheric pressure

90

15