Description of features – Amico Hydraulic Patient Transfer Stretcher (S-H-300) User Manual
Page 7
Head Section Articulation
!
WARNING:
Surface positions that cause the patient’s torso to be at an angle less than 90°
to the legs could reduce circulatory efficiency in the lower extremities. Such
positions are not recommended for extended periods of time and should be
under medical supervision. Patient injury could occur.
!
WARNING:
Make sure to fully control the lift mechanism when there is little or no weight
on it. Failure to do so could cause it to rise quickly and cause injury or equipment damage.
To raise the Head Section:
Squeeze the bar under the head section and lift up on the head section.
Release the bar when the desired position is reached. The head section locks in place.
To lower the Head Section:
Squeeze the bar under the head section and push down on the head section.
Release the bar when the desired position is reached. The head section locks in place.
Knee Surface Articulation Crank
To raise the Knee Surface:
a. Pull out the crank handle from its stored position under the
foot section.
b. Turn the crank handle clockwise until the knee surface reaches the
desired position.
To lower the Knee Surface:
c. Pull out the crank handle from its stored position under the foot section.
d. Turn the crank handle counterclockwise until the knee surface reaches the desired position.
After raising or lowering the knee surface, return the crank to its stored position.
Description of Features
Permanent Telescopic IV Pole
The permanent IV pole is available as an option on the S-H-TRAUMA stretcher.
!
WARNING:
Do not exceed the load capacity of the IV pole. If the IV pole is overloaded,
personal injury or equipment damage may occur.
The safe working load of the pole is 25 lbs.
To Raise the permanent IV pole:
a. Grasp the stored IV pole and raise it from its stored position. Make sure it is secure in its support bracket, with
the sleeve over the support bracket.
b. Grasp the top section of the IV pole, and raise it until the desired height is reached. Screw back knob to hold
the IV pole in position.
To lower the permanent IV pole:
a. Unscrew the knob to release the IV pole.
b. Manually lower the top section of the pole until it fully collapses.
c. Screw back knob to hold the IV pole in position.
Removable Utility Tray
To remove utility tray, lift gently upwards to ensure poles are free from the
bushing housing. To attach utility tray, align poles with bushing housing and
press gently downwards.
X-RAY Cassette Holder
To position Cassette holder at desired location of patient’s body,
turn crank handle clockwise or counter clockwise as needed.