beautypg.com

System overview, I g-dualscan, Instrumentation room chamber – ORBIT FR G-DualScan User Manual

Page 3

background image

128

G-DualScan uses a Vector Network Analyzer as the
RF source/receiver for antenna measurements. The
Amplification Unit has RF amplifiers for each of the RX
and TX channels. G-DualScan uses a Transfer Switching
Unit to emit from the AUT to the probe(s) or vice versa.
A dedicated RF switch allows the selection of either the
single-probe or the multi-probe set-up. The Positioner

Controller drives the goniometer and azimuth axes for the
AUT, and the elevation axis for the gantry arm. Measure-
ments can be performed in both continuous wave and
optional pulsed mode. In the case of phased array anten-
na measurement, the system utilizes an optional real time
controller to control and synchronize the measurement
system with the device under test.

1

4

3

2

Real Time

Controller

Positioner

Controller

INSTRUMENTATION ROOM

CHAMBER

Rx

Tx

Triggers

Triggers

USB

GPIB

Primary

Synthesizer

Data Acquisition

& Processing Platform

Auxiliary

Synthesizer

Mixer Unit

Amplification

Unit

Transfer

Switching Unit

RF

Switch

N-PAC

5

System overview

3

I G-DualScan