System overview, I g-dualscan, Instrumentation room chamber – ORBIT FR G-DualScan User Manual
Page 3
128
G-DualScan uses a Vector Network Analyzer as the
RF source/receiver for antenna measurements. The
Amplification Unit has RF amplifiers for each of the RX
and TX channels. G-DualScan uses a Transfer Switching
Unit to emit from the AUT to the probe(s) or vice versa.
A dedicated RF switch allows the selection of either the
single-probe or the multi-probe set-up. The Positioner
Controller drives the goniometer and azimuth axes for the
AUT, and the elevation axis for the gantry arm. Measure-
ments can be performed in both continuous wave and
optional pulsed mode. In the case of phased array anten-
na measurement, the system utilizes an optional real time
controller to control and synchronize the measurement
system with the device under test.
1
4
3
2
Real Time
Controller
Positioner
Controller
INSTRUMENTATION ROOM
CHAMBER
Rx
Tx
Triggers
Triggers
USB
GPIB
Primary
Synthesizer
Data Acquisition
& Processing Platform
Auxiliary
Synthesizer
Mixer Unit
Amplification
Unit
Transfer
Switching Unit
RF
Switch
N-PAC
5
System overview
3
I G-DualScan