VICI Selectors (multiposition) Microelectric User Manual

Page 7

background image

7

Serial Commands

NP

Displays the number of positions the actuator is currently set to index

NPnn

Sets the number of positions (nn) for the current valve

CWnn

Sends the actuator clockwise to position nn (from 1 to NP)

CCnn

Sends the actuator counterclockwise to position nn (from 1 to NP)

GOnn

Sends the actuator to position nn (from 1 to NP) via the shortest route

CP

Displays the current position

SD

Displays the digital input status where 0 = enabled and 1 = disabled

SDn

Sets the digital input status to [0] BCD, [1] disabled, [2] parallel, or

[3] binary. NOTE: Setting the status to [1] locks out communication

through the manual controller and remote control cables, so digital input

status defaults to [0] at power up.

SM

Displays the current default rotational direction for the digital inputs.

SMl

Sets the default rotational direction for the digital inputs to [F] for forward

rotation, [R] for reverse rotation, or [A] to automatically choose the

shortest route. This feature can be used instead of external wiring tied

to the Manual and/or Autodirection input lines on the Digital I/O port.

ID

Displays the current device ID setting

IDn

Sets the device ID to value n, from 0 to 9

NOTE: When the ID feature is enabled, all commands to the device

must be prefaced by the ID. Entering ID* clears the ID.

SB

Displays the current baud rate

SBnnnn

Sets the baud rate to 1200, 2400, 4800, 9600 (default), 14400,

19200, 28800, or 38400. The parity setting, number of data bits,

and number of stop bits cannot be changed. (See section entitled

Setting a New Baud Rate on the next page)

SL

Displays current Data Latch signal status

SLn

This command displays or changes the requirement for a Data Latch

signal to accompany BCD inputs. When set to [0] (factory default), the

data latch is required for BCD inputs. When set to [1], the data latch is

NOT required. This feature can reduce the number of control lines

required for a system with a dedicated BCD output port and only one

actuator connected. NOTE: Be sure all the BCD inputs are asserted

within 20 milliseconds of each other or the actuator may be misdirected.

SO

Displays the current offset value

SOnn

Sets the offset value of the first position to be any number from 1 to

96 minus the total number of positions. Example: for a 10 position

valve, the offset can be set from 1 to 86. (See section entitled Using

the Offset Feature beginning on the next page.)

VR

Displays the part number and date of the firmware

/?

Displays a list of valid commands

Note: In this chart, nn represents numbers to be entered and l represents letters.