Detector Fault Determination
Electrical faults in the detector are most easily discovered by making
resistance measurements between pins of the detector cable connector. The
table below shows the pin numbering and connections for the detector cable
(view into pins). Nominal resistance values (in ohms) at 25°C are shown.
Unshaded boxed without numbers represent open circuits, and should in no
case be blow 1 M
Ω.
Resistance values given should be within ±10% of the
nominal values.
Pin
1
2
3
4
5
6
8
9
11
12
13
2
3
4
0
5
25
25
6
108
8
9
0
11 12
25
25
13
0
14
50