Standby and shutdown procedures, Factory repair service, Standby – VICI 120 Dynacalibrator User Manual
Page 9: Shutdown
Standby and Shutdown Procedures
Standby
1. When the Dynacalibrator is not in use, switch the mode control selector
to the STANDBY position. This will divert all flow through the overflow
vent and scrubber to atmosphere. The scrubber will filter the permeant
gas so that no concentration of the permeant will escape out into envi-
ronment.
2. Turn the flow control knob counterclockwise to set the pump at a mini-
mum flow setting
3. Leave the permeation device in the permeation chamber to maintain
equilibrium until you are ready to use the Model 120 again.
Shutdown
1. Remove the permeation device from the chamber and place it in the
shipping tube with the charcoal packets. If the device will not be used for
over 1 week and its total useful life is under 1 year, it is recommended
that the device be placed in cold storage to prolong its useful life.
2. Recharge the battery if required.
Factory Repair Service
VICI provides complete repair and calibration services for Dynacalibrators.
If service is required, contact the facility indicated below for authorization
and repair specifics prior to returning your unit.
North America, South America, and
Australia/Oceania
Europe, Asia, and Africa
VICI Metronics Inc.
VICI AG International
tel: (877) 737-1887
fax: (360) 697-6682
[email protected]
tel: Int + 41 41 925-6200
fax: Int + 41 41 925-6201
[email protected]
Equipment should be shipped in the original or equivalent packing materials,
prepaid and insured. Include a complete written description of symptoms,
problems, or calibration requirements, as well as the name and phone
number of the person to contact for discussion of the required service.
Upon receipt, VICI will inspect the equipment and advise the contact
person of any unusual repair time or cost factors prior to starting the
work. Repaired equipment will be returned with an invoice for any
costs not covered by warranty.
Remove all permeation tubes from the chamber before
packaging and shipping the Dynacalibrator