Sample calculation – VICI 120 Dynacalibrator User Manual

Page 8

background image

Calculating the Concentration

The following flow correction factor table can be used if the elevation is
known. Find the elevation and temperature of your operational environment
on this table and multiply this factor to the published calibrated flow. The
calibrated flow is included in the certification documents shipped with the
instrument.

Temperature (°C)

Elevation

0

5

10

15

20

25

30

35

40

-1000'

1.064

1.054

1.045

1.036

1.027

1.018

1.010

1.002

0.994

sea level

1.045

1.035

1.026

1.017

1.009

1.000

0.992

0.984

0.976

1000'

1.026

1.017

1.008

0.999

0.990

0.982

0.974

0.966

0.958

2000'

1.007

0.998

0.989

0.981

0.972

0.964

0.956

0.948

0.941

3000'

0.989

0.980

0.971

0.963

0.955

0.947

0.939

0.931

0.924

4000'

0.971

0.962

0.954

0.945

0.937

0.929

0.922

0.914

0.907

5000'

0.953

0.944

0.936

0.928

0.920

0.912

0.905

0.897

0.890

Figure 3: Flow correction factor table

Sample Calculation

Given:
Permeation rate: 21,000 ng/min Cl

2

@ 30°C

Flow: 1100 cc/min (from Flowmeter Table
Ambient temperature: 15°C
Ambient pressure: (unknown), elevation = 2000

'

Correct the flow per table above:
F = (0.981)(1100) = 1079 cc/min

Then:

��������������������� �����

���������������

����

���������

For zero reference measurement, remove the permeation device from the
chamber.