Detector electrical connections, Initial power-up – VICI D-4-I-SH14-R User Manual
Page 14

10
Detector Electrical Connections
CAUTION: Do not use a wrench to tighten the SMC
connectors on the bias and electrometer cables.
Connections should be finger tight only.
1. Referring to Figures 5 and 6 as necessary, connect the BIAS cable to
the top electrode and the electrometer cable (ELECT) to the bottom
electrode.
2. Connect the high-voltage cable from the detector to the pulse supply.
Initial Power-Up
CAUTION: Always make sure that discharge gas is
flowing before heating and powering up the detector.
1. Before installing the column, set the gas flow to 30 ml/min (measured at
the detector vent). Let it flow for 15 minutes so that all air is purged from
the helium purifier.
2. Plug in the helium purifier and turn on the GC.
PULSER
MODULE
ELECTROMETER CABLE
P/N: I-23230-SH14
BIAS CABLE
P/N: I-23404-SH14
SHIMADZU
ELECTROMETER
BOARD
TO 110/220 VAC
PULSER
POWER
SUPPLY
Figure 6: Schematic of electrical connections
Installation