beautypg.com

Troubleshooting, No peaks, Low sensitivity – VICI D-4-I Quick Start User Manual

Page 3: High background current, High noise level

background image

Troubleshooting

No Peaks

If the background current is stable but there are no peaks:

1. Check column flow.

2. Check the column insertion position (11.4 cm for

capillary columns, 8.9 cm for packed columns).

3. If there are still no peaks, check to make sure that

the GC is actually making an injection.

4. If there are still no peaks, consult Valco.

Low Sensitivity

High Background Current

High Noise Level

If the noise level is high:

1. See if it improves with the GC fan turned off.

2. Check the column insertion position (11.4 cm for

capillary columns, 8.9 cm for packed columns).

3. If the noise level is still high, do a hydrogen leak test

(described on the next page).

4. If there are no leaks found, or leaks are found and

repaired and the noise is still high, consult Valco.

Purple/

blue

Pink

Disconnect the column
and cap the column inlet

Baseline current is still high
(> 2.5 nA)

Leaks in the fittings of the
PDD plumbing

No detectable leaks.

Do a hydrogen leak test
(described on last page)

Check
discharge
color

Leaks inside the detector
cell itself

Consult

Valco

Tighten leaking fittings

The detector isn’t leaking.
Check for column bleed
and/or leaks in the GC setup.

Baseline current falls to
normal range (

£

2.5 nA)

Possible bad carrier gas
or empty He bottle. Install
a new bottle.

Consult

Valco

Consult

Valco

Consult

Valco

Baseline current is still high
(> 2.5 nA)

Baseline current falls to
normal range (

£

2.5 nA)

Baseline current is still high
(> 2.5 nA)

Baseline current falls to
normal range (

£

2.5 nA)

Baseline current is still high
(> 2.5 nA)

Baseline current falls to
normal range (

£

2.5 nA)

Sensitivity is in
acceptable range

Sensitivity is still
too low

Sensitivity is in
acceptable range

Sensitivity is still
too low

Sensitivity is in
acceptable range

Sensitivity is still
too low

Sensitivity is in
acceptable range

Remove column and
condition it, then run
a sample

Background current
remains high

Background current
drops to normal range

Check column inlet for
leaks. Tighten fittings.

Background current
increases

Check column position
according to drawing on
page 2, and run a sample.

Background current
remains in normal range

Remove column and
bake detector at
300°C for four hours.
Install column and run
a sample.

Run
a sample

Refer to “Troubleshooting:
High Background Current”

Reinstall column and check
background current

Check background current
with column removed and
column inlet plugged

Normal background current
(

£

2.5 nA)

High background current
(

>

2.5 nA)

Consult

Valco

Sensitivity is still
too low