beautypg.com

7 disassembly – Trebor 620D Magnum User Manual

Page 16

background image

PAGE 16

MAGNUM 620D PUMP OPERATION / MAINTENANCE MANUAL

4.7 DISASSEMBLY

During the life of the pump it will be necessary to perform certain preventive
maintenance procedures to ensure its continued high performance. This section
and the next (4.8

, Assembly) are provided for the user’s convenience in

disassembly and re-assembly procedures. Reference the exploded view
drawing and parts list for assistance in identifying components.

Immerse or flush the pump assembly using DI water and a neutralizing

agent.

Install mounting fittings in pump adapter ports and lock body into bench

mounting fixture. NOTE: Securely attach mounting fixture to work surface
using hardware provided.

Figure 4-1

Remove transfer tubes and fittings from pump head.

Figure 4-2

Using a strap wrench, turn heads counter-clockwise to remove.

Remove head and check diaphragms for cracks or cuts.