Symptom, Cause, 5 diagnostic chart – HF scientific CLX Online Residual Chlorine Monitor User Manual
Page 32: 6 technical and customer assistance

CLX (1/12)
Page 27
REV 5.1
9.5 Diagnostic Chart
Symptom
Cause
Cure
Lower display shows MA
4-20 mA loop open
Check wiring. See sections
4.3.4 and 7.2
Lower display shows FAIL
Major system fault
Refer to section 9.1
Readings are erratic
(1) Bubbles in solution
(2) Debris in flow
(1) See above
(2) Install T strainer at inlet
Readings are lower than expected (1) Condensate or leaky
measurement cuvette
(2) Measurement cuvette
dirty
(3) Reagents bad or expired
(4) Buffer reagent not being
dispensed
(1) Install desiccant cartridge
kit
(2) Replace or clean cuvette
(3) Replace reagents
(4) Check buffer lines and
check valves.
Upper display flashes
Sample Over-Range
Check sample. Sample may be
too high to read.
Upper display shows nOnE while
attempting to calibrate
No current reading displayed
Wait for CLX to post a reading