beautypg.com

Gossen Mavolux 5032C/B User Manual

Page 8

background image

8

GOSSEN Foto- und Lichtmeßtechnik GmbH

Characteristics MAVOLUX 5032B

Most Important Error Limits MAVOLUX 5032B

Light Detector

silicon photodiode with
V(

λ

) filter per DIN 5032 part 7,

class B

Error Limits

per DIN 5032, part 7, class B

Meas. Quantity

Measuring Range

Resolution
in lx

Resolution
in fc

in Lux (lx)

in footcandles (fc)

Illumination

I

0.01

19,99

0.001

1.999

0.01

0.001

I I

0.1

1

99,9

0.01

19.99

0.1

0.01

I I I

1

19

99

0,1

1

99.9

1

0.1

I V

10

19 990

1

19

99

10

1

in Candela/m² (cd/m²)

in footlambert (fL)

in cd/m²

in fL

Luminance
(with luminance
attachment
for cd/m

2

or fL)

I

0.1

199.9

0.01

19.99

0.1

0.01

I I

1

1

999

0.1

199.9

1

0.1

I I I

10

19

990

1

1

999

10

1

I V

100

199 900

10

19 990

100

10

Characteristics

Allowable Error
Limits per
DIN 5032 Class B

Actual Error
MAVOLUX 5032B

V(

λ

) Matching

6%

3 %

True Cosine Evaluation

3 %

2 %

Linearity

2 %

1 %

Adjustment Error

1 %

0,8%

Total Error

10 %