Gossen Mavolux 5032C/B User Manual
Page 8

8
GOSSEN Foto- und Lichtmeßtechnik GmbH
Characteristics MAVOLUX 5032B
Most Important Error Limits MAVOLUX 5032B
Light Detector
silicon photodiode with
V(
λ
) filter per DIN 5032 part 7,
class B
Error Limits
per DIN 5032, part 7, class B
Meas. Quantity
Measuring Range
Resolution
in lx
Resolution
in fc
in Lux (lx)
in footcandles (fc)
Illumination
I
0.01
…
19,99
0.001
…
1.999
0.01
0.001
I I
0.1
…
1
99,9
0.01
…
19.99
0.1
0.01
I I I
1
…
19
99
0,1
…
1
99.9
1
0.1
I V
10
…
19 990
1
…
19
99
10
1
in Candela/m² (cd/m²)
in footlambert (fL)
in cd/m²
in fL
Luminance
(with luminance
attachment
for cd/m
2
or fL)
I
0.1
…
199.9
0.01
…
19.99
0.1
0.01
I I
1
…
1
999
0.1
…
199.9
1
0.1
I I I
10
…
19
990
1
…
1
999
10
1
I V
100
…
199 900
10
…
19 990
100
10
Characteristics
Allowable Error
Limits per
DIN 5032 Class B
Actual Error
MAVOLUX 5032B
V(
λ
) Matching
6%
3 %
True Cosine Evaluation
3 %
2 %
Linearity
2 %
1 %
Adjustment Error
1 %
0,8%
Total Error
10 %
8 %