2 installation – Metrohm Flow Cell for Bioscan – 6.5331.0X0 User Manual
Page 9

■■■■■■■■■■■■■■■■■■■■■■
2 Installation
IC Equipment Set Flow Cell for Bioscan
■■■■■■■■
3
2 Installation
Caution
Never switch on the flow cell when it is..
■
..not being rinsed through with a conducting eluent at the time; or
■
..not completely connected; or
■
..moist on the outside so that a short circuit could occur between
the connections of working eleoctrode and auxiliary electrode.
8
1
2
3
4
5
6
7
Figure 1
Parts and connections of flow cell
1
Reference electrode connection
For black electrode cable (6.2156.000).
2
Inlet for eluent
Connection of PEEK capillary by 6.2744.130
KELF pressure screw only. PEEK pressure
screws may damage the threads of the cell.
3
Auxiliary electrode connection
For blue electrode cable (6.2156.000).
4
Working eleoctrode connection
For red electrode cable (6.2156.000).
5
Outlet for eluent
Connection of PEEK capillary by 6.2744.130
KELF pressure screw only. PEEK pressure
screws may damage the threads of the cell.
6
Mounting screws (4x)
7
Distance piece 50
µm (6.1254.020)
Defines the volume of the cell.
8
Check marks
For correct alignment of the top and bottom
part of the flow-through cell.