2 Installation

IC Equipment Set Flow Cell for Bioscan

■■■■■■■■

3

2 Installation

Caution

Never switch on the flow cell when it is..

..not being rinsed through with a conducting eluent at the time; or

..not completely connected; or

..moist on the outside so that a short circuit could occur between
the connections of working eleoctrode and auxiliary electrode.

8

1

2

3

4

5

6

7

Figure 1

Parts and connections of flow cell

1

Reference electrode connection
For black electrode cable (6.2156.000).

2

Inlet for eluent
Connection of PEEK capillary by 6.2744.130
KELF pressure screw only. PEEK pressure
screws may damage the threads of the cell.

3

Auxiliary electrode connection
For blue electrode cable (6.2156.000).

4

Working eleoctrode connection
For red electrode cable (6.2156.000).

5

Outlet for eluent
Connection of PEEK capillary by 6.2744.130
KELF pressure screw only. PEEK pressure
screws may damage the threads of the cell.

6

Mounting screws (4x)

7

Distance piece 50

µm (6.1254.020)

Defines the volume of the cell.

8

Check marks
For correct alignment of the top and bottom
part of the flow-through cell.