3 Installation

IC-Ausrüstung: MiPT

■■■■■■■■

9

3.3

Equipping the supply bottle

Required accessories

Bottle (6.1608.070) filled with ultrapure water

Eluent bottle cap (6.1602.160)

Adsorber tube (6.1619.000), included in the accessories for the dosing
unit (6.3032.120)

Adapter for adsorber tube (6.1624.000)

FEP aspiration tubing (6.1819.110)

M8 stopper (6.1446.080), included in the accessories for the eluent
bottle cap (6.1602.160)

1 Mounting the aspiration tubing

Insert the aspiration tubing into the M6 opening of the eluent
bottle cap.

Use the capillary cutter to cut the aspiration tubing to such a
length that it touches the bottom of the bottle.

2 Inserting the stopper

Tighten the M8 stopper in the M8 opening of the eluent bottle
cap.

3 Mounting the adsorber tube

Fill the adsorber tube with some cotton and adsorber material.

Place the adsorber tube on the adapter.

Insert the adapter into the SGJ opening of the eluent bottle cap.