3 equipping the supply bottle, Equipping the supply bottle – Metrohm MiPT User Manual
Page 15
■■■■■■■■■■■■■■■■■■■■■■
3 Installation
IC-Ausrüstung: MiPT
■■■■■■■■
9
3.3
Equipping the supply bottle
Required accessories
■
Bottle (6.1608.070) filled with ultrapure water
■
Eluent bottle cap (6.1602.160)
■
Adsorber tube (6.1619.000), included in the accessories for the dosing
unit (6.3032.120)
■
Adapter for adsorber tube (6.1624.000)
■
FEP aspiration tubing (6.1819.110)
■
M8 stopper (6.1446.080), included in the accessories for the eluent
bottle cap (6.1602.160)
1 Mounting the aspiration tubing
■
Insert the aspiration tubing into the M6 opening of the eluent
bottle cap.
■
Use the capillary cutter to cut the aspiration tubing to such a
length that it touches the bottom of the bottle.
2 Inserting the stopper
■
Tighten the M8 stopper in the M8 opening of the eluent bottle
cap.
3 Mounting the adsorber tube
■
Fill the adsorber tube with some cotton and adsorber material.
■
Place the adsorber tube on the adapter.
■
Insert the adapter into the SGJ opening of the eluent bottle cap.