4 ambient temperature, 5 capacity, 6 placement – Metrohm Power Box (for 899 Coulometer) User Manual
Page 21: Ambient temperature, Capacity, Placement

■■■■■■■■■■■■■■■■■■■■■■
5 Technical specifications
Power Box
■■■■■■■■
15
■
EN/IEC 61000-4-6
■
EN/IEC 61000-4-11
■
EN/IEC 61000-4-14
■
EN/IEC 61000-4-28
5.4
Ambient temperature
Nominal function
range
0 - +40 °C
(at a maximum of 85% humidity)
Storage
–20 - +60 °C
Transport
–20 - +60 °C
5.5
Capacity
Maximum
capacity
4 Ah
Usable capacity
90%
Work with a 899
Coulometer:
12 h of continuous operation during standard titration (120 s of titra-
tion, 180 s of conditioning, 1200 µg of water).
Remarks
After a typical titration operation lasting 8 h, the Power Box can be
charged for approximately 1 h. This can allow for another 8 h of oper-
ation. Afterwards, the Power Box must be charged for 14 h (full
charge).
5.6
Placement
Safety distance
Minimum safety distance between the rear panel of the Power Box and
parts and devices that can generate sparks:
■
10 cm (in accordance with EN 50272-2)
An 899 Coulometer can be placed on top of the Power Box.